420 Plus Installation Manual

2.3.1

Cable Grounding

Except for the power cord, all cables routed through 
the cord grips should be grounded against the 
indicator enclosure. Do the following to ground 
shielded cables:

Use the lockwashers, clamps, and kep nuts 
provided in the parts kit to install grounding 
clamps on the enclosure studs adjacent to cord 
grips. Install grounding clamps only for cord 
grips that will be used; do not tighten nuts.

Route cables through cord grips and 
grounding clamps to determine cable lengths 
required to reach cable connectors. Mark 
cables to remove insulation and shield as 
described below:

For cables with foil shielding, strip insulation 
and foil from the cable half an inch (15 mm) 
past the grounding clamp (see Figure 2-2). 
Fold the foil shield back on the cable where 
the cable passes through the clamp. Ensure 
silver (conductive) side of foil is turned 
outward for contact with the grounding 
clamp.

For cables with braided shielding, strip cable 
insulation and braided shield from a point just 
past the grounding clamp. Strip another half 
inch (15 mm) of insulation 

only

 to expose the 

braid where the cable passes through the 
clamp (see Figure 2-2).

For load cell cables, cut the shield wire just 
past th e gro unding  clamp . Shield wire 
function is provided by contact between the 
cable shield and the grounding clamp.

Route stripped cables through cord grips and 
clamps. Ensure shields contact grounding 
clamps as shown in Figure 2-2. Tighten 
grounding clamp nuts.

Finish installation using cable mounts and ties 
to secure cables inside of indicator enclosure.

Figure 2-2. Grounding Clamp Attachment for Foil-Shielded 

and Braided Cabling

2.3.2

Bypass Power Button

If jumper JMP1 is set for SW position, the power 
switch on the front panel of the indicator will turn the 
unit ON or OFF. If it is not in the SW position, the 
indicator will power up as soon as the AC is applied. 
This allows the front panel overlay power control 
switch to be bypassed. Label “F” in Figure 2-3 shows 
the location of  JMP1.  Figure 2-4 on page 7 shows 
entire board.

Figure 2-3. 420 Plus CPU and Power Supply Board JMP1 

Shunt Location

Insulated cable 

Foil  (silver side out) 

Grounding clamp 

Shield wir e  (cut) 

Length of foil before folding 
back on cable insulation  

Cut insulation here 
for foil-shielded cables 

Braid 

Cut insulation here 
for braided cables 

NOTE:

 Install lockwashers 

first, against backplate, 
under grounding clamp 

Содержание 420 Plus

Страница 1: ...85127 420 Plus HMI Digital Weight Indicator Version 1 14 Installation Manual...

Страница 2: ......

Страница 3: ...ed Tare Value 3 1 4 7 Acquire Parts Sample 4 1 4 8 Display Part Weight 4 1 4 9 Display Accumulator 4 1 4 10 Display or Change Time 4 1 4 11 Display or Change Setpoint Value 4 1 4 12 Turn Setpoint On o...

Страница 4: ...5 2 Saving and Transferring Data 38 5 2 1 Saving Indicator Data to a Personal Computer 38 5 2 2 Downloading Configuration Data from PC to Indicator 38 6 0 Print Formatting 39 6 1 Print Formatting Com...

Страница 5: ...20 4 20 mA tracking of gross or net weight values Available in 115 VAC and 230 VAC versions Available in DC power version The 420 Plus is NTEP certified and pending Measurement Canada approval for Cl...

Страница 6: ...Gross weight is within 0 25 graduations of zero This annunciator lights when the scale is zeroed Standstill Scale is at standstill or within the specified motion band Some operations including tare fu...

Страница 7: ...tween primary and secondary units The appropriate units LED to the right of the display is lit 1 4 3 Zero Scale 1 In gross mode remove all weight from the scale and wait for the standstill annunciator...

Страница 8: ...date press the TIME DATE key once press TIME DATE a second time to display the time To set the date press the TIME DATE key once Use the numeric keypad to enter the date then press the ENTER key Use t...

Страница 9: ...N 42149 Three reducing glands PN 15664 One capacity label PN 42350 Three each of grounding clamps PN 53075 external tooth lock washers PN 15133 and kep nuts PN 14626 for cable shield grounding against...

Страница 10: ...se the braid where the cable passes through the clamp see Figure 2 2 For load cell cables cut the shield wire just past the grounding clamp Shield wire function is provided by contact between the cabl...

Страница 11: ...420 Plus Installation Manual Installation 7 Figure 2 4 420 Plus CPU and Power Supply Board...

Страница 12: ...runs voltage drop over the power conductor needs to be considered See table and formula below to compute voltage drop VDROP 2 85A x 1000ft length of run in ft 2 85A maximum current draw from DC DC po...

Страница 13: ...board see Figure 2 4 Use cable ties to secure serial cables to the inside of the enclosure The EDP port supports full duplex RS 232 communications only the serial port provides either active 20 mA ou...

Страница 14: ...kplate over the enclosure and reinstall the backplate screws Use the torque pattern shown in Figure 2 6 to prevent distorting the backplate gasket Torque screws to 15 in lb 1 7 N m Figure 2 6 420 Plus...

Страница 15: ...lt stand 1 11 44676 Sealing washer 1 12 42640 Screws 1 4 28NF X 1 4 2 13 19538 Cable grip plugs 2 14 15626 Cable grips 3 15 30375 Nylon seal rings for cable grips 3 16 85202 Power cord assembly 115 VA...

Страница 16: ...lation Manual Figure 2 8 420 Plus Overlay and Power Supply 20 21 22 To J7 header on PC Board 6 From Power Cord 16 23 28 5 pin overlay 8 pin overlay Ground Wire Detail 25 27 24 To backplate 3 From powe...

Страница 17: ...420 Plus Installation Manual Installation 13 Figure 2 9 420 Plus Enclosure Backplate and CPU Board 1 2 3 4 5 6 7 8 9 10 12 11 13 14 15 16 17 18 32 30...

Страница 18: ...ay Revolution supports both uploading and downloading of indicator configuration data This capability allows configuration data to be retrieved from one indicator edited then downloaded to another To...

Страница 19: ...d overload tare function sample rate and digital filtering parameters FORMAT Format Set format of primary and secondary units display rate CALIBR Calibration Calibrate indicator See Section 4 0 on pag...

Страница 20: ...ad see Figure 3 4 Figure 3 4 Editing Procedure for Numeric Values 3 2 Menu Structures and Parameter Descriptions The following sections provide graphic representations of the 420 Plus menu structures...

Страница 21: ...gal value varies depending on local regulations For scales using linear calibration do not set the zero tracking band to a value greater than that specified for the first linearization point ZRANGE 1...

Страница 22: ...immediate area of the scale Choices indicate the number of A D conversions that are averaged to obtain the displayed reading A higher number gives a more accurate display by minimizing the effect of a...

Страница 23: ...u parameter descriptions DSPRAT 250MS 500MS 750MS 1SEC 1 5SEC 2SEC 2 5SEC 3SEC 4SEC 6SEC 8SEC Display rate Sets the update rate for displayed values Values are in milliseconds MS or seconds SEC Table...

Страница 24: ...88 88 Decimal point location Determines the location of the decimal point or dummy zeros in the secondary unit display DSPDIV 5D 1D 2D Display divisions Selects the value of minimum division size of t...

Страница 25: ...ivolt calibration press ENTER to display and edit the test weight value Press ENTER again to display and edit the millivolt value for that weight Press ENTER a third time to calibrate and display the...

Страница 26: ...TE INDUST LFT Specifies stream rate Stream rate can be set to industrial or legal for trade PRNDES EDP PRN BOTH Print destination Selects the EDP port printer port or both ports simultaneously for dat...

Страница 27: ...rameter Choices Description Table 3 5 Serial Menu Parameters Continued Note XXXXXXX XXXXXXX PWRUPM GO OIML REGULA 000000 number CONSNU DELAY SETPNT PROGRM PFORMT SERIAL CALIBR CONFIG FORMAT 000000 num...

Страница 28: ...value Specifies the initial consecutive number CONSNU value used when the consecutive number is reset by sending CLRCN digital input UID 1 Specifies the unit identification number any numeric value up...

Страница 29: ...ORMAT Display and edit active character and ASCII value Delete active character Scroll left in formatting string Increment ASCII value of active character Decrement ASCII value of active character Pre...

Страница 30: ...an the setpoint value or is within or outside of the band value LOWER means the output is active until you reach weight If trip is HIGHER the output is active when the setpoint is met or exceeded INBA...

Страница 31: ...resets the consecutive number to the value specified on the CONSTU parameter PROGRM menu KBDLOC disables the keypad while the digital input is held low HOLD holds the current display Releasing this in...

Страница 32: ...20 mA output Select 0 for 0 10 V output select 20 for 4 20 mA output ERRACT FULLSC HOLD ZEROSC Error action Specifies how the analog output responds to system error conditions Possible values are FULL...

Страница 33: ...The VERS menu is used to check the software version and reg version installed in the indicator You can also check the indicator model Figure 3 14 Version Menu XXXXXXX ALGOUT XXXXXXX XXXXXXX Software v...

Страница 34: ...he test weights to be used for the span calibration Press to store the entered WVAL value and go to span calibration WSPAN 5 With WSPAN displayed place test weights on the scale and press to calibrate...

Страница 35: ...rate zero The indicator displays CAL while calibration is in progress 3 Place test weights on the scale and use the WVAL command to enter the test weight value in the following format WVAL nnnnnn CR 4...

Страница 36: ...ficients 4 4 1 Adjusting Final Calibration Calibration may be affected by environmental factors including wind vibration and angular loading For example if the scale is calibrated with 1000 lb a strai...

Страница 37: ...used in both setup and weighing mode Several of the commands serve as pseudo keys providing functions that are not represented by a key on the front panel For example to enter a 15 pound tare weight...

Страница 38: ...r the equal sign If you type an incorrect command or value the display reads Changes to the parameters are saved as they are entered but typically do not take effect until you exit setup mode For exam...

Страница 39: ...on REZERO Rezero LC CD Set deadload coefficient value LC CW Set span coefficient value Table 5 5 CALIBR EDP Commands Command Description Values EDP BAUD EDP port baud rate 300 600 1200 2400 4800 9600...

Страница 40: ...ription Values GFMT Gross demand print format string See Section 6 0 on page 39 for detailed information NFMT Net demand print format string CFMT Count demand print format string Table 5 8 PFORMT EDP...

Страница 41: ...mmand Description Response Format UID Set unit ID nnnnnnn SD Set date MMDDYY DDMMYY YYMMDD enter using DATEFMT specified ST Set time hhmm enter using 24 hour format SX Start EDP streaming OK or EX Sto...

Страница 42: ...parameters to the PC as ASCII formatted text 5 2 2 Downloading Configuration Data from PC to Indicator Configuration data saved on a PC or floppy disk can be downloaded from the PC to an indicator Th...

Страница 43: ...each print format string includes the output field length of the print formatting commands not the command length For example if the indicator is configured to show a decimal point the G command gener...

Страница 44: ...is format might look like the following FINE TRANSFER CO 32400 WEST HIGHWAY ROAD SMALLTOWN 1345 lb GROSS 6 2 2 Using the Front Panel If you have no access to equipment for communication through the ED...

Страница 45: ...he print format without the formatting commands NL and SP required by the front panel or EDP command methods Using Revolution you can type text directly into the grid then select weight value fields f...

Страница 46: ...Printer format checksum error Call RLWS Service LCCKSM Load cell calibration checksum error Recalibrate load cells EIDATA Internal RAM checksum error Call RLWS Service E REF A D reference error A D c...

Страница 47: ...y query which annunciators are currently displayed on the indicator front panel The ZZ command returns the currently displayed weight and a decimal number representing the LED annunciators currently l...

Страница 48: ...Figure 7 1 Continuous Output Data Format STX POL wwwwwww UNIT G N S TERM ASCII 02 decimal Weight data 7 digits right justi ed with decimal point leading zero supression Overload Underrange Display ove...

Страница 49: ...CK 06 06 38 26 F 70 46 f 102 66 Ctrl G BEL 07 07 39 27 G 71 47 g 103 67 Ctrl H BS 08 08 40 28 H 72 48 h 104 68 Ctrl I HT 09 09 41 29 I 73 49 i 105 69 Ctrl J LF 10 0A 42 2A J 74 4A j 106 6A Ctrl K VT 1...

Страница 50: ...CB 235 EB 140 8C 1 4 172 AC 204 CC 236 EC 141 8D 173 AD 205 CD 237 ED 142 8E 174 AE 206 CE 238 EE 143 8F 175 AF 207 CF 239 EF 144 90 176 B0 208 D0 240 F0 145 91 177 B1 209 D1 241 F1 146 92 178 B2 210...

Страница 51: ...anel Display Characters Figure 7 2 shows the 7 segment LED character set used to display alphanumeric characters on the 420 Plus front panel Figure 7 2 420 Plus Display Characters 9 E Q F R G S 0 H T...

Страница 52: ...s and the secondary decimal point is set to 8888 88 the indicator will overflow if 5 tons or more are applied to the scale With 5 tons applied and a conversion factor of 2000 the secondary units displ...

Страница 53: ...tering effect is 20 4 8 8 With this configuration each A D reading has a 1 in 20 effect on the displayed weight value Setting the filters to 1 effectively disables digital filtering 7 7 2 DFSENS and D...

Страница 54: ...cause more consecutive out of band readings so DFSENS should be set higher to counter low frequency transients Reconfigure as necessary to find the lowest effective value for the DFSENS parameter 7 8...

Страница 55: ...the front panel key functions in test mode Note that because the Test Menu functions are all on a single menu level the GROSS NET key has no function Press the ZERO key to exit test mode Table 7 7 sum...

Страница 56: ...emory Test Green Pass Red Fail LED ANA Red Not Implemented Analog Option Card Digital Input Output Status DO 1 Digital Out 1 On when lit active low DO2 Digital Out 2 On when lit active low DI 1 Digita...

Страница 57: ...erload 12 V continuous static discharge protected RFI Protection Signal excitation and sense lines protected by capacitor bypass Analog Output Optional fully isolated voltage or current output 14 bit...

Страница 58: ...t caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equipment has not been modifi...

Страница 59: ......

Страница 60: ...PN 85127 05 12...

Отзывы: