250-10-2 Bariatric Handrail Scale
12
Visit our website
ASTART
Weight process start limit — Maximum (full
capacity)/10; Determine when the weight
algorithm starts (when the “----” is dis
-
played), below this value the scale will
show live weight
Value (2.0)
The display toggles between a numeric value and
ASTART
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from ASTART to ARW; If you want to change the
value, use the following steps;
1.Press
ENTER
key to change value starting with the right most digit.
2.Use the Up/Down arrow keys to increment/decrement numbers.
3.Press
BMI
key to move to the left.
4.Use the Up/Down arrow keys to increment/decrement numbers.
5.When done, press
ENTER
key to move to the next parameter (ARW).
ARW
Auto Reweigh — Restarts the weight algo
-
rithm if the weight changed by more than
this value.
Value (4.0 lb)
The display toggles between a numeric value and
ARW
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from ARW to SAL; If you want to change the value,
use the following steps;
1.Press
ENTER
key to change value starting with the right most digit.
2.Use the Up/Down arrow keys to increment/decrement numbers.
3.Press
BMI
key to move to the left.
4.Use the Up/Down arrow keys to increment/decrement numbers.
5.When done, press
ENTER
key to move to the next parameter (SAL).
SAL
Semi Auto Live —This value is the interval
between weight displays during the algo
-
rithm process
Value (0.5)
The display toggles between a numeric value and
SAL
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from SAL to ROUND; If you want to change the
value, use the following steps;
1.Press
ENTER
key to change value starting with the right most digit.
2.Use the Up/Down arrow keys to increment/decrement numbers.
3.When done, press
ENTER
key to move to the next parameter (ROUND).
ROUND
Scale Resolution —
Values in kg:
1
, 2, 5, 10, 20, 50, 100
Values in lb: 1,
2
, 5, 10, 20, 50, 100, 200
0.2
0.5
1.0
2.0
5.0
10.0
20.0
0.1
The display toggles between a numeric value and
ROUND
;
The decimal point location is set to the DISP parameter display decimal
point location;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from ROUND to DISP; If you want to change the
value, use the following steps;
1.Press
ENTER
key to change value.
2.Press the Up/Down arrow keys to change the available parameters.
3.When done, press
ENTER
key to move to the next parameter (DISP).
DISP
0.0
0
0.0000
0.000
0.00
The display toggles between a numeric value and
DISP
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from DISP to BAUD; If you want to change the
value, use the following steps;
1.Press
ENTER
key to change value.
2.Use the Up/Down arrow keys to change the available parameters.
3.When done, press
ENTER
key to move to the next parameter (BAUD).
BAUD
Baud rate
96
48
1152
576
384
288
192
144
Indicator display illustrates first two digits of baud rate only;
The display toggles between a numeric value and baud;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from BAUD to ATOL;
If you want to change the value, use the following steps
1.Press
ENTER
key to change value.
2.Use the Up/Down arrow keys to change the available parameters.
3.When done, press
ENTER
key to move to the next parameter (ATOL).
ATOL
Algorithm initial tolerance — Maximum
value is 255.
Values above 255 will not let you proceed
and will return to the previous value.
Value (10)
The display toggles between a numeric value and
ATOL
;
If you do not want to change this value, press the
BMI
key to move to the
next setting; Example: from ATOL to ALEN; If you want to change the
value, use the following steps;
1.Press
ENTER
key to change value starting with the right most digit.
2.Use the Up/Down arrow keys to increment/decrement numbers.
3.Press the
BMI
key to move to the left.
4.When done, press
ENTER
key to move to the next parameter (ALEN).
Parameter
Description
Choices
Steps
Table 4-1. Configuration Mode Menu (Continued)
Содержание 250-10-2
Страница 26: ...250 10 2 Bariatric Handrail Scale 22 Visit our website www RiceLake com...
Страница 27: ......