250-10-2 Bariatric Handrail Scale
6
Visit our website
3.2 Weighing
Use the following steps to weigh.
1. Press
to turn on the scale.
0.0
appears on the display along with the
ZERO
annunciator.
2. Place the patient on the scale. The patient’s weight is displayed, the
LOCK
annunciator is on and the indicator beeps
to indicate the end of the weighing process.
3. Press
to change the display from lb to kg and vice-versa.
4. Press and hold
until
OFF
displays to turn off the scale.
3.3 Hold/Release Function
Use the following steps to use the Hold/Release function.
1. Press
to turn on the scale.
0.0
prompts along with
ZERO
on the display.
2. Press
once the patient’s weight stabilizes. The patient’s weight and the
HOLD
and
LOCK
annunciators remain
on the display when the patient is off the scale.
3. Press
again to return the scale to zero.
Pressing
will not work while using Hold/Release function.
Pressing
prior to the patient getting on the scale will
also hold the weight display.
3.4 Preset Tare
Use the following steps for the Preset Tare function prior to patient weighing if additional items are being used by the patient.
1. Press
to turn on the scale.
0.0
appears on the display along with the
ZERO
annunciator.
2. Place additional item(s)
on the scale.
3. Press
until the display returns to
0.0
and
NET
annunciator appears on the display.
4. Remove additional item(s) from the scale. The weight displays with a negative symbol to the left of it.
Not removing the additional item(s) prior to patient weighing will also work.
5. Position the patient and additional item(s) on the scale. The display identifies the patient weight. The
NET
annunciator
is still active. The weight of the additional item(s) remains stored in memory for the duration of this weigh in.
6. To cancel the tare weight, remove patient from the scale and press
until
NET
disappears from the display and
the display turns back to
0.0
and
GROSS
appears.
Tare weight is also canceled when the scale is turned off.
3.5 Toggle Tare
Use the following steps to use the Toggle Tare function when the additional item to be weighed is known.
1. Press
when the scale is empty and
0.0
displays. The default values prompts while
0.0
is flashing on the display
(default is programmed to be 33.0 lb/15.0 kg).
2. Use
and
to adjust the value. Press
to start the tare function. The
NET
annunciator turns on instead
of the
GROSS
annunciator.
Note
Note
TARE
T
Note
TARE
T
Note
TARE
T
ENTER
Содержание 250-10-2
Страница 26: ...250 10 2 Bariatric Handrail Scale 22 Visit our website www RiceLake com...
Страница 27: ......