 Digital Physician Operation Instructions

5.0

Troubleshooting and Testing

Refer to the following instructions to check and correct any failure before contacting service personnel.

Table 5-1.  Troubleshooting Table for the Rice Lake Scale Line

Symptom

Possible Cause

Corrective Action

Scale does not turn on

Dead battery

Connect the scale to a power source.

Faulty electrical outlet

Use a different electrical outlet.

Bad power supply

Replace adaptor.

Questionable weight or the scale 
does not zero

External object is interfering with the 
scale

Remove the interfering object from the scale.

Display did not show 0.0 before 
weighing

Help the patient off the scale, zero the scale and 
begin the weighing process again.

Scale is not placed on a level floor

Ensure the scale is level and begin the weighing 
process again.

Scale is out of calibration

Check the weight with a known weight value.

Improper tare

Place the tare item on the scale. Press 

REWEIGH

. Once the weight of the item is 

displayed, press 

TARE

. Place the patient back on 

the scale. Press the 

REWEIGH

 button again.

Weighing is performed but the display 
shows 

WEIGH

 and 

REWEIGH

 every 

few seconds; the weighing process 
takes too long and no weight is 
displayed

The patient is not sitting still

Ask the patient to be still.

The display shows a 

STOP

 message

The load on the scale exceeds the 
capacity of the scale 

Remove the excess weight and use the scale 
according to manufacture’s specs.

The display shows 

LO Bat

 message

The battery is low

Recharge the battery.

The display shows Err message as 
detailed in the table below

Err 2

Low saturation state (low A/D)

The load cell is not connected properly. Check 
the cables and mechanical connections. If the 
problem persists, replace the set of load cells.

Err 3

High saturation state (high A/D)

See Err 2

Err 6

Unstable weight. Cannot calibrate

Check the load cells’ mechanical surroundings 
and see that nothing touches them and that the 
cables are properly welded.

Содержание 140-10 Series

Страница 1: ...113684 Rice Lake Digital Physician Scale 140 10 Series Software Revision 11387 Operation Instructions To be the best by every measure...

Страница 2: ......

Страница 3: ...d Testing 8 6 0 Maintenance 9 6 1 Basic Maintenance 9 6 2 Cleaning 9 7 0 Digital Physician Scale Specifications 10 For More Information 11 Digital Physician Scale Limited Warranty 12 Technical trainin...

Страница 4: ...ii Digital Physician Operation Instructions Rice Lake continually offers web based video training on a growing selection of product related topics at no cost Visit www ricelake com webinars...

Страница 5: ...so designed with a large hook handle and two heavy duty rear wheels for easy portability This manual can be viewed and downloaded from the Rice Lake Weighing Systems web site at www ricelake com healt...

Страница 6: ...er scale removal and set up 2 Carefully lift the scale out of the packaging material that it came in lifting it out by the scale base NOTE DO NOT lift the scale out of the box by just the scale column...

Страница 7: ...approximately 75 hours If the LO Bat indicator is showing on the display recharge the battery or connect the scale to an AC power source as soon as possible for accurate weighing Battery Charging When...

Страница 8: ...very sensitive so only a gentle pushing motion is required to obtain results The scales have the capability of performing different operations beyond just calculating weight The various operating ins...

Страница 9: ...ard Weight Status BMI Weight Status Below 18 5 Underweight 18 5 24 9 Normal 25 0 29 9 Overweight 30 0 and Above Obese The following examples show weight ranges the corresponding BMI ranges and the wei...

Страница 10: ...ey until TARE disappears from the display The tare weight is also cancelled when the scale is turned off Use the following steps to enter a tare without placing that item on the scale An example of th...

Страница 11: ...t With a stable in range weight press and hold the Kg Lb Print key for at least three seconds or until the scale emits two quick beeps Note that if the scale does not beep after five seconds then rele...

Страница 12: ...ht value Improper tare Place the tare item on the scale Press REWEIGH Once the weight of the item is displayed press TARE Place the patient back on the scale Press the REWEIGH button again Weighing is...

Страница 13: ...gh the following steps for basic maintenance 1 Check the overall appearance of the entire scale for any obvious signs of damage abuse etc 2 Inspect the condition of the AC adaptor for cord cracking or...

Страница 14: ...nmental Operating Temperature 50 to 104 F 14 to 40 C Storage Temperature 32 to 158 F 0 to 70 C Humidity 85 relative humidity Capacity and Graduation Digital Physician Scale 550lb 250kg 0 2lb 100g Cert...

Страница 15: ...n Saturday CST Telephone Sales Technical Support 800 472 6703 Canadian and Mexican Customers 800 321 6703 International 715 234 9171 Immediate Emergency Service For immediate assistance call toll free...

Страница 16: ...s and was not caused by accident misuse neglect alteration improper installation improper repair or improper testing RLWS shall be the sole judge of all alleged non conformities Such equipment has not...

Страница 17: ......

Страница 18: ...PN 113684 10 10...

Отзывы: