5.7.3
Malfunctions
WARNING situations, which can be controlled by the User, are being indicated as shown below.
If the recommended actions have been taken, press the key .
After the fifth indication (door open or lack of chemical) the machine will switch over into FAULT
condition.
See warnings from the table below:
Warning text
Explanation
Chemical tank empty
Change or re-fill the detergent container.
Close door
The loading door is open or not closed
correctly
Chamber too hot for starting the cycle.
Wait or open the door for cooling the
chamber. Or press cancel to stop the cycle.
A new cycle will not start if the chamber
temperature is over 60
0
C
No disinfection
Must be confirmed by the user if the cycle
does not include disinfection phase.
Cycle OK, No disinfection
Must be confirmed by the user if the cycle
does not include disinfection phase.
Open water valves
Water valves shall be open when starting.
Power cut during the cycle
There has been a power cut and
the ongoing cycle must be restarted
Parameters are not OK
Contact service / Deko distributor
Banner warnings:
Banner warnings appear as scrolling text at the top of the user screens and
in the warning log of service display.
See maintenance manual APPENDIX 1.