602400011500
Revision C
Page
17
of
24
NOTE: Once the “SetClock.txt” file has been
created, it is important to follow the steps below
as quickly as possible to ensure a minimum RTC
time error exists.
6.3.5
Save the “SetClock.txt” file on the USB
device, and then move the USB device to
the dehydrator’s front panel USB port.
6.3.6
To update the RTC, cycle power on the
dehydrator using the power cord.
6.3.7
The dehydrator’s RUN TIME indicator will
flash as a log file is written to the USB
device. Then the RTC date and time will
be updated.
6.3.8
The dehydrator’s LOW PRESSURE
indicator will flash 5 times (once per
second) to indicate successful RTC
update.
6.3.9
The unit will then start normally.
6.4
Replacing
the
Lid
&
Placing
the
Unit
Back
In
Service:
6.4.1
Refer to
Picture
6.1
and
Picture
6.2
.
Place the lid next to the unit, fan down.
Connect the case fan connector to the
control board’s fan header, wires facing
in.
6.4.2
Pivot the lid onto the unit, making sure
not to pull or bind the fan wires.
6.4.3
Replace the lid screws. Insert all loosely at
first, and then torque each screw to 10
inch-pounds (1.13 Newton-meters).
6.4.4
Replace the unit in its original installation
location. If present, reattach the alarm
port connector. If present, reconnect the
site ground stud. Reconnect the 3/8” air
tube to the unit’s output fitting on the
rear panel. (See section 5.6).
6.4.5
Open any distribution system valves that
were closed when the unit was removed.
6.4.6
Plug the unit in and verify that it starts
and runs normally. Verify that the case
fan is operating correctly.