General Information
Safe Place Transmitters User Guide
Page 15 of 36
0510-1121-G
WARNING
: Only Smart Sense banding
material can be used to fasten Smart Sense
transmitters. Use of any other banding
material with the Smart Sense transmitters
voids the RF Technologies' Product
Warranty and could result in system failure.
2. With the yellow-side of the banding
material facing outwards, use the plastic
guide tab to thread the banding material
from the bottom slot out through the top
slot on the first clamp.
CAUTION:
Do not thread both sides of the
banding material through the clamps until
you are ready to secure the transmitter to
the infant. Pre-threading may allow the
transmitter to auto-enroll and alarm.
3. Pull the tab until it clears the clamp and
push the clamp lever down until it snaps
shut.
4. Place the transmitter with banding
material around the infant’s ankle and
thread the banding material through the
second transmitter clamp slot.
CAUTION
: Ensure transmitter is banded
properly. It should fit snugly against the skin
without an air gap between the transmitter
and skin, yet allow the band and transmitter
to be turned on the ankle. Over-tightening
the band will not reduce the number of
alarms and could cause patient injury.