RF-LAMBDA USA www.rflambda.com
Sales: [email protected] Technical : [email protected]
W
ide B
an
d E
M
C B
en
ch
top
Pow
er A
mp
lifier
18
GH
z ~
40
GH
z
REMC18G40GC
Rev 1. 02-02-2021
Ensure that the amplifier input and output ports are safely terminated into a proper 50 ohm load before turning on the power.
Never operate the amplifier without a load. A proper 50 ohm load is defined as a load with impedance less than 1.9:1 or return
loss larger than 10dB relative to 50 Ohm within the specified operating band width.
Power Supply Requirements
Power supply must be able to provide adequate current for the amplifier. Power supply should be able to provide 1.5 times the
typical current or 1.2 times the maximum current (whichever is greater).
In most cases, RF - Lambda amplifiers will withstand severe mismatches without damage. However, operation with poor loads
is discouraged. If prolonged operation with poor or unknown loads is expected, an external device such as an isolator or
circulator should be used to protect the amplifier.
Ensure that the power is off when connecting or disconnecting the input or output of the amp.
Prevent overdriving the amplifier. Do not exceed the recommended input power level.
Adequate heat-sinking required for RF amplifier modules. Please inquire.
Amplifiers do not contain Thermal protection, Reverse DC polarity or Over voltage protection with the exception of a few
models. Please inquire.
Proper electrostatic discharge (ESD) precautions are recommended to avoid performance degradation or loss of functionality.
What is not covered with warranty?
Each RF - Lambda amplifier will go through power and temperature stress testing.
Since the die, ICs or MMICs are fragile, these are not covered by warranty. Any damage to these will NOT be free to repair.
Amplifier Use
Part No.
Description
REMC18G40GC
18-40GHz Power Amplifier
Ordering Information