3.7 Clear Measurements From Memory

If you are sure that you want to permanently remove all stored memories, press the 

SET

 button 7 times until  

“ ” appears when the power is off. Press the 

START/STOP

 button until “ ” flashes 3 times to clear the 

memory. After this, press the 

MEM

 button, the symbols “ ” and “

” will appear on the screen. This means 

that the memory has successfully cleared.

NOTE: 

You cannot clear individual measurements.

3.8 Discontinuing a Measurement 

If it is necessary to interrupt a blood pressure measurement for any reason (e.g. the patient feels unwell), the 

START/

STOP

 button can be pressed at any time. The device then immediately lowers the cuff pressure automatically.

4. Error Messages / Malfunctions

If an error occurs during a measurement the LCD displays the corresponding error code.

Error

Possible Cause

Remedy

Weak signal or a sudden change in pressure

Ensure that the cuff is worn correctly 

and measure again. Avoid movement or 

talking when the cuff is being inflated.

External disturbance (eg. if there is a cell 

phone near by)

Electromagnetic interference can cause 

odd behaviour in the monitor. Move away 

from the source.

Appears when there is an error during the 

inflation process

Make sure the cuff is wrapped around 

the arm properly and that the air plug is 

properly connected to the unit.

Abnormal blood pressure

Rest for 5 minutes and try measurement 

again.

Low battery

Replace batteries.

If problems occur when using the device the following points should be checked, and if necessary, the 

corresponding measures should be taken.

Malfunction

Remedy

The display remains blank when the device is 

switched on.

1.  Check batteries for correct polarity.

2.  If using the AC adapter, ensure it is correctly 

inserted.

The pressure does not rise even though the pump 

is running.

Check the connection of the cuff tube and connect 

properly if necessary.

Содержание RX810

Страница 1: ...Model RX810 Instruction Manual 10 17 18 ...

Страница 2: ...Screen 2 3 Inserting the Batteries 2 4 Using the AC DC Power Adapter 3 Using the Device 3 1 Select the User 3 2 Setting the Time Date and Year 3 3 Obtaining Accurate Measurements 3 3ATips on Taking Accurate Measurements 3 3B Common Sources of Errors 3 3C Fitting the Cuff 3 4 Measuring Your Blood Pressure 3 5 Hypertension Classification Indicator 3 6 Viewing Previously Recorded Values 3 7 Clear Mea...

Страница 3: ...ortant Information Refer to the following sections to learn about important safety instructions and how to take care of the RexallTM MC Premium Blood Pressure Monitor 1 2A Safety Information Self measurement means control not diagnosis or treatment Your values must always be discussed with your doctor or a physician who is familiar with your family history If you are undergoing medical treatment a...

Страница 4: ...e display is not suitable for checking the frequency of heart pacemakers If you have been diagnosed with a severe arrhythmia or irregular heartbeat vascular constriction liver disorder or diabetes have a cardiac pacemaker or are pregnant measurements made with this instrument should only be evaluated after consultation with your doctor Take care while handling the batteries in the device Incorrect...

Страница 5: ...nts The most accurate way to compare devices is to take two readings at the same time However most people and doctor s offices do not have the equipment necessary to measure blood pressure from two devices simultaneously To take sequential measurements properly requires a pair of initial measurements to determine the subjects blood pressure level first with the reference equipment followed by 60 s...

Страница 6: ...re We differentiate between common primary essential hypertension and secondary hypertension The latter group can be attributed to specific organic malfunctions Please consult your doctor for information about the possible origins of your own increased blood pressure values 1 4 Normal Blood Pressure Values Blood pressure is too high when measuring at home and you have rested the diastolic pressure...

Страница 7: ...r blood pressure to fluctuate significantly 50 mmHg to 60 mmHg throughout the day Blood pressure is normally lowest at night but increases during waking hours when the stress and activities of everyday life are highest mmHg AM Time PM 40 50 60 70 80 90 100 110 120 130 140 150 160 170 8 00 10 00 12 00 Typical 24 Hours Rhythm 14 00 16 00 18 00 20 00 22 00 24 00 2 00 4 00 6 00 8 00 systolic blood pre...

Страница 8: ...tarted 2 1 About the Blood Pressure Monitor RX810 This section describes the various components of the monitor Cuff Socket SET Button Memory Button Wide Range Cuff Cuff Connector AC DC Adapter Socket START STOP Button Item Function LCD Screen Recorded blood pressure heart rate and time are displayed here AC DC Socket Connect to adapter Memory Button Press to view the previously recorded data or to...

Страница 9: ...rect polarity as indicated 3 Replace the battery compartment cover Attention After the battery warning appears the device is blocked until the batteries have been replaced Please use AA Long Life or Alkaline 1 5V batteries If the blood pressure monitor is left unused for long periods please remove the batteries from the device Do not mix old and new batteries Do not mix alkaline standard carbon zi...

Страница 10: ... using the MEM buttons Finish the process by pressing the SET button and the USER YEAR MONTH DATE HOUR and MINUTES will all be stored in the products memory NOTE In order to change any setting you must repeat the process and confirm each setting by pushing the SET button 3 3 Obtaining Accurate Measurements Your blood pressure can vary based on numerous factors physiological conditions and your sur...

Страница 11: ...lood to flow away 3 3C Fitting the Cuff a Pass the end of the cuff through the flat metal ring so that a loop is formed The closure must be facing outward Ignore this step if the cuff has already been prepared b Wrap the cuff around your bare upper left arm The rubber tube should be on the inside of your arm extending downward to your hand c Lay the cuff on the arm as illustrated Make certain that...

Страница 12: ...s been appropriately positioned the measurement can begin 1 Press the START STOP button until all the symbols appear on the screen after will flash for 2 seconds The pump then begins to inflate the cuff The increasing pressure in the cuff is shown on the display 2 After the suitable pressure has been reached the pump stops and the cuff gradually deflates The cuff pressure will then be displayed In...

Страница 13: ...ur measurement has five bars it is in the red Stage 2 Hypertension zone If your measurement has six bars it is in the red Stage 3 Hypertension zone Red Red Red Yelllow Green Green Red Red Red Yelllow Green Green Red Red Red Yelllow Green Green 3 6 Viewing Previously Recorded Values The blood pressure monitor automatically stores your measurements with time and date It can store up to 90 measuremen...

Страница 14: ... signal or a sudden change in pressure Ensure that the cuff is worn correctly and measure again Avoid movement or talking when the cuff is being inflated External disturbance eg if there is a cell phone near by Electromagnetic interference can cause odd behaviour in the monitor Move away from the source Appears when there is an error during the inflation process Make sure the cuff is wrapped aroun...

Страница 15: ...ight bubble Handle this carefully and avoid all types of stress through twisting or buckling c Clean the device with a soft dry cloth Do not use gas thinners or similar solvents Spots on the cuff can be removed carefully with a damp cloth and soapsuds The cuff with bladder must not be washed in a dishwasher clothes washer or submerged in water d Handle the tube carefully Do not pull on it Do not a...

Страница 16: ...return shipping 7 Technical Specifications Weight 250g with batteries Size 100mm W x 132mm L x 45mm H Storage temperature 20 C to 55 C 4 F to 131 F Humidity 10 to 85 relative humidity maximum Operation temperature 10 C to 40 C 50 F to 104 F Display LCD Display Liquid Crystal Display Measuring method Oscillometric Pressure sensor Capacitive Measuring range SYS DIA 30 to 299 mmHg Pulse 40 to 199 bea...

Страница 17: ...tructions and keep the instruction manual for future use Type BF applied part Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations not with domestic waste IP20 Protected against solid foreign particles with a diameter of more than 12 5 mm no protection against water ...

Отзывы: