R E V O L U T I O N A R Y   S C I E N C E  

Saniclave RS

-

SC

-

200

 

5. 

Low Water:  

The display will read 

'Err 5'

 if there is an insufficient  

 

 

 

amount of water in the chamber. Press SET to  

 

 

 

acknowledge the error. Fill reservoir up to the tempera 

 

 

 

ture sensing probe. Restart cycle. 

 
6.    

Exhaust tank full: 

After every five cycles, the Saniclave display will flash 

 

 

 

'

drn tank full

.'

 

 At this point, the exhaust tank must be 

 

 

 

emptied if the end of the drain tube is not already rest

 -

 

 

 

ing in a sink or bucket.

 

 Revolutionary Science recom

 

 

 

 

mends that the drain tube be left in the sink or other 

 

 

 

 

reservoir for regular drainage.

 

 The drain tank warning 

 

 

 

can be bypassed by pressing any button.

 

 If exhaust 

 

 

 

 

tank is not drained after every sixth cycle, water may 

 

 

 

 

overflow. 

 

7.  Door is difficult to open or close:  

 

 

 

There may be a few reasons why the door is difficult to 

 

 

 

open or close.   

 

 

 

1) Ensure that the door seal is tucked under all lugs in the 

 

 

 

door.  (see picture on page 19) 

 

 

 

2) The gasket offset spring must rest against the outside 

 

 

 

diameter of the seal and should not overlap the seal.  

 

 

 

 3) Check to make sure door markings line up properly.  

 

 

 

Misaligned door markings will inhibit the door functionally 

 

 

 

4)  The chamber may have not yet depressurized.   

  

 

 

Wait up to five minutes and then attempt to open the  

 

 

 

door.   

 

 

 

5) the door stop may be caught on the door clip near the 

 

 

 

bottom of the door.  This can be dislodged by pushing up 

 

 

 

on the bottom of the plastic door housing. 

 

 

 

6) The door seal may need lubricant.  To do this, remove 

 

 

 

and dry the seal completely.  lightly dust the seal with  

 

 

 

talcum powder and spread evenly around the entire seal 

 

 

 

(or talc based baby powder) approximately once every 50 

 

 

 

to 100 cycles or as needed.  

Caution:  Too much  

 

 

 

lubrication may compromise the seal.

  

8.  Display reads '8888' 

 

 

 

This means that your real time clock has reset.  The time 

 

 

 

and date settings will reset if the Saniclave is left un 

 

 

 

plugged longer than three days.  See page 9 for setting the 

 

 

 

date and time.  Unplug and replug the power cord.  The 

 

 

 

display should now read '0000' and you can commence  

 

 

 

with sterilization.  The real time clock does not need to 

 

 

 

be set properly in order to operate the Saniclave. 

 

 

Содержание saniclave 200 RS-SC-200

Страница 1: ......

Страница 2: ...SC 200 tray and support rack in correct...

Страница 3: ...ed Use 4 Safeguards and Cautions 5 Operating the Saniclave 200 6 Printer 6 Troubleshooting and Error Modes 10 Warranty Registration and Policy 12 Specifications 13 Contact Information 13 Checklist and...

Страница 4: ...that the Saniclave 200 will provide years of reliable service The high quality of construction and durable parts deliver a performance equivalent to much costlier equipment You are invited to visit ou...

Страница 5: ...Saniclave RS SC 200 RE VO LU TIO N A RY SC IE N CE 3 Display Symbols Control and Parts...

Страница 6: ...1 257 0633 3 Carefully read this instruction manual before operating the Saniclave 200 4 Fill out warranty registration card and mail it to Revolutionary Science 17319 Lake Blvd Shafer MN 55074 Intend...

Страница 7: ...may be added to the water to increase detecta bility Class III distilled water is recommended 12 NEVER reuse water left in the exhaust tank Re used water may contain endotoxins Drain water tank and r...

Страница 8: ...rmophilus at least weekly to confirm proper sterilization For additional information on purchasing biological indicators see page 12 Printer installation and operation Saniclave 200 The recommended pr...

Страница 9: ...d with the CLOSE marking on the autoclave Never attempt to rotate the door past this CLOSE marking Improper closing of the door can cause pressure leaks or damage to the autoclave 8 Press the door in...

Страница 10: ...reads Fill if the START button is pressed and the chamber has not yet been filled with sufficient water Preheat Mode Display will read Ph for Preheat This mode will usually last for about 30 minutes...

Страница 11: ...e reset If set correctly the time and date will be printed on the printout Print After a cycle is complete when display reads CC press the SET button once Other than the printouts the Saniclave 200 do...

Страница 12: ...g up markings on door and on body and twisting the door closed Restart the cycle 3 Over Temperature Alarm If temperature exceeds the maximum of 124 C the display will read Err 3 Allow unit to drain th...

Страница 13: ...must rest against the outside diameter of the seal and should not overlap the seal 3 Check to make sure door markings line up properly Misaligned door markings will inhibit the door functionally 4 The...

Страница 14: ...______ Intended use _______________________________________________________________ Where did you hear of our product _______________________________________________________________ Comments _________...

Страница 15: ...d product for medical use K112811 FDA Registered Revolutionary Science is an FDA Registered Company CE IEC Safety RS SC 200 tested to and found to be compliant with IEC 61010A and associated safety st...

Страница 16: ...________________ Your Notes _______________________________________________________________ _______________________________________________________________ ____________________________________________...

Страница 17: ...to printer manufacturer s instructions For regular Sterilizer Monitoring Guidelines see page 16 When storing the Saniclave 200 1 Disconnect Saniclave 200 from mains 2 Drain all water from the chamber...

Страница 18: ...rator Cycle Display reading at cycle end Signature 102281A 10 45 00 AM 11 21 12 121C for 30 minutes Parameters were met 1 CC _________ _______ 121C for 30 minutes _________ _______ 121C for 30 minutes...

Страница 19: ...cycle parameters of the Saniclave 200 When monitoring the cycle with such indicators load the chamber according to your facility s standard operating procedures Chemical Indicators The purpose of chem...

Страница 20: ...hall be used to monitor sterility of loads processed in Saniclave 200 Class 5 Integrators A Class 5 integrator are a type of chemical indicator It does not contain microbial organisms A Class 5 Integr...

Страница 21: ...19 R E V O L U T I O N A R Y S C I E N C E Saniclave RS SC 200 Proper placement of the door seal...

Страница 22: ......

Отзывы: