Saniclave RS

-

SC

-

200

 

Troubleshooting

 

Error Conditions and Display Readings    

Note: 

If a failure occurs, the load must be considered non-sterile.

  The 

sterilizer and load should then be investigated.  Contact Revolutionary Science if 
problem persists. 

1. Manual Cycle Abort: 

 

If 'START' is pressed, display will read '

Err  1

' for 

 

 

 

 

cycle abort.  User may abort the cycle anytime, 

 

 

 

 

during, preheat, sterilize or dry modes.  If cycle 

 

 

 

 

is aborted, the contents of the load shall be  

 

 

 

 

considered non-sterile.  Press SET to             

 

 

 

 

acknowledge the error.  

 

2. Preheat too long (Door Ajar): 

If door is not closed properly and the cycle is  

 

 

 

 

initiated, the cycle will fail when the autoclave  

 

 

 

 

cannot pressurize.  Display will read '

Err 2

.'  

 

 

 

 

Press SET to acknowledge the error. Take off  

 

 

 

 

the door. Remove door gasket. Inspect the door 

 

 

 

 

gasket for cracks.  Replace with a new gasket if 

 

 

 

 

necessary.  Reinstall  gasket  into  door.    Properly 

 

 

 

 

close door by lining up markings on door and on 

 

 

 

 

body and twisting  the door closed.  Restart the 

 

 

 

 

cycle. 

3. 

Over Temperature Alarm:

 

If temperature exceeds the maximum of 124°C 

 

 

 

 

the display will read '

Err  3

.'  Allow unit to drain 

 

 

 

 

the reservoir fully into the exhaust tank or sink. 

 

 

 

 

Press SET to acknowledge the error.  Open  

 

 

 

 

door  when  chamber  has  cooled.  Remove  load 

 

 

 

 

from chamber. Refill reservoir. Reinsert load.  

 

 

 

 

Close the door. Restart cycle.  If an   over  

 

 

 

 

temperature condition occurs again, contact  

 

 

 

 

Revolutionary Science to investigate the  

 

 

 

 

problem further.  Discontinue use until problem 

 

 

 

 

is resolved. 

4. 

Under Temperature Alarm:

  If the temperature or pressure of the   

 

 

 

 

chamber falls below sterilization temperature of 

 

 

 

 

121°C  during  the  sterilization  cycle,  the  digital 

 

 

 

 

display will read '

Err 4

.'  Press SET to  

 

 

 

 

acknowledge the error.  Open door when  

 

 

 

 

chamber has cooled. Remove load from  

 

 

 

 

chamber.  Refill  reservoir

Reinsert  load.  Close 

 

 

 

 

the door. Restart  cycle.  If an under  

 

 

 

 

temperature condition occurs again, contact  

 

 

 

 

Revolutionary Science to investigate further.  

 

 

 

 

Discontinue use until problem is resolved. 

 

R E V O L U T I O N A R Y   S C I E N C E  

Содержание saniclave 200 RS-SC-200

Страница 1: ......

Страница 2: ...SC 200 tray and support rack in correct...

Страница 3: ...ed Use 4 Safeguards and Cautions 5 Operating the Saniclave 200 6 Printer 6 Troubleshooting and Error Modes 10 Warranty Registration and Policy 12 Specifications 13 Contact Information 13 Checklist and...

Страница 4: ...that the Saniclave 200 will provide years of reliable service The high quality of construction and durable parts deliver a performance equivalent to much costlier equipment You are invited to visit ou...

Страница 5: ...Saniclave RS SC 200 RE VO LU TIO N A RY SC IE N CE 3 Display Symbols Control and Parts...

Страница 6: ...1 257 0633 3 Carefully read this instruction manual before operating the Saniclave 200 4 Fill out warranty registration card and mail it to Revolutionary Science 17319 Lake Blvd Shafer MN 55074 Intend...

Страница 7: ...may be added to the water to increase detecta bility Class III distilled water is recommended 12 NEVER reuse water left in the exhaust tank Re used water may contain endotoxins Drain water tank and r...

Страница 8: ...rmophilus at least weekly to confirm proper sterilization For additional information on purchasing biological indicators see page 12 Printer installation and operation Saniclave 200 The recommended pr...

Страница 9: ...d with the CLOSE marking on the autoclave Never attempt to rotate the door past this CLOSE marking Improper closing of the door can cause pressure leaks or damage to the autoclave 8 Press the door in...

Страница 10: ...reads Fill if the START button is pressed and the chamber has not yet been filled with sufficient water Preheat Mode Display will read Ph for Preheat This mode will usually last for about 30 minutes...

Страница 11: ...e reset If set correctly the time and date will be printed on the printout Print After a cycle is complete when display reads CC press the SET button once Other than the printouts the Saniclave 200 do...

Страница 12: ...g up markings on door and on body and twisting the door closed Restart the cycle 3 Over Temperature Alarm If temperature exceeds the maximum of 124 C the display will read Err 3 Allow unit to drain th...

Страница 13: ...must rest against the outside diameter of the seal and should not overlap the seal 3 Check to make sure door markings line up properly Misaligned door markings will inhibit the door functionally 4 The...

Страница 14: ...______ Intended use _______________________________________________________________ Where did you hear of our product _______________________________________________________________ Comments _________...

Страница 15: ...d product for medical use K112811 FDA Registered Revolutionary Science is an FDA Registered Company CE IEC Safety RS SC 200 tested to and found to be compliant with IEC 61010A and associated safety st...

Страница 16: ...________________ Your Notes _______________________________________________________________ _______________________________________________________________ ____________________________________________...

Страница 17: ...to printer manufacturer s instructions For regular Sterilizer Monitoring Guidelines see page 16 When storing the Saniclave 200 1 Disconnect Saniclave 200 from mains 2 Drain all water from the chamber...

Страница 18: ...rator Cycle Display reading at cycle end Signature 102281A 10 45 00 AM 11 21 12 121C for 30 minutes Parameters were met 1 CC _________ _______ 121C for 30 minutes _________ _______ 121C for 30 minutes...

Страница 19: ...cycle parameters of the Saniclave 200 When monitoring the cycle with such indicators load the chamber according to your facility s standard operating procedures Chemical Indicators The purpose of chem...

Страница 20: ...hall be used to monitor sterility of loads processed in Saniclave 200 Class 5 Integrators A Class 5 integrator are a type of chemical indicator It does not contain microbial organisms A Class 5 Integr...

Страница 21: ...19 R E V O L U T I O N A R Y S C I E N C E Saniclave RS SC 200 Proper placement of the door seal...

Страница 22: ......

Отзывы: