Troubleshooting
Error Conditions and Display Readings
Note:
If a failure occurs, the load must be considered non-sterile.
The
sterilizer and load should then be investigated. Contact Revolutionary Science if
problem persists.
1. Manual Cycle Abort:
If 'START' is pressed, display will read '
Err 1
' for
cycle abort. User may abort the cycle anytime,
during, preheat, sterilize or dry modes. If cycle
is aborted, the contents of the load shall be
considered non-sterile. Press SET to
acknowledge the error.
2. Preheat too long (Door Ajar):
If door is not closed properly and the cycle is
initiated, the cycle will fail when the autoclave
cannot pressurize. Display will read '
Err 2
.'
Press SET to acknowledge the error. Take off
the door. Remove door gasket. Inspect the door
gasket for cracks. Replace with a new gasket if
necessary. Reinstall gasket into door. Properly
close door by lining up markings on door and on
body and twisting the door closed. Restart the
cycle.
3.
Over Temperature Alarm:
If temperature exceeds the maximum of 124°C
the display will read '
Err 3
.' Allow unit to drain
the reservoir fully into the exhaust tank or sink.
Press SET to acknowledge the error. Open
door when chamber has cooled. Remove load
from chamber. Refill reservoir. Reinsert load.
Close the door. Restart cycle. If an over
temperature condition occurs again, contact
Revolutionary Science to investigate the
problem further. Discontinue use until problem
is resolved.
4.
Under Temperature Alarm:
If the temperature or pressure of the
chamber falls below sterilization temperature of
121°C during the sterilization cycle, the digital
display will read '
Err 4
.' Press SET to
acknowledge the error. Open door when
chamber has cooled. Remove load from
chamber. Refill reservoir
.
Reinsert load. Close
the door. Restart cycle. If an under
temperature condition occurs again, contact
Revolutionary Science to investigate further.
Discontinue use until problem is resolved.
R E V O L U T I O N A R Y S C I E N C E
Содержание saniclave 200 RS-SC-200
Страница 1: ......
Страница 2: ...SC 200 tray and support rack in correct...
Страница 5: ...Saniclave RS SC 200 RE VO LU TIO N A RY SC IE N CE 3 Display Symbols Control and Parts...
Страница 21: ...19 R E V O L U T I O N A R Y S C I E N C E Saniclave RS SC 200 Proper placement of the door seal...
Страница 22: ......