Preparation for Sterilization (Default mode)
1.
Warm up cycle
: Run a warm up cycle each day prior to use. Cycle may be aborted during the
sterilization mode to conserve on time. Remove door. Unplug the cord, replug the cord and
commence with sterilization needs.
2.
Preclean instruments according to their manufacturer's recommendations.
3. Follow instrument manufacturer’s instructions for correct cleaning and lubricating procedures.
WARNING! Failure to properly clean instruments before autoclaving, may result in unsterile
equipment. Using unsterile equipment may result in serious injury or death.
4. It is recommended that all instruments be dismantled, if possible.
5. Make sure packs do not touch. Do not stack packs or single piece instruments on top of each
other.
CAUTION: Stacking instruments improperly may result in unsterile equipment!
6. Use a sterilization indicator strip with each set of instruments being autoclaved.
7.
The recommended load is 4.5 lbs (not including the rack).
Never exceed the maximum load
.
Exceeding maximum load may result in unsterile equipment.
8. Run a biological indicator spore test (Geobacillus stearothermophilus) at least weekly to confirm
proper sterilization. For additional information on purchasing biological indicators, see page 12.
Printer installation and operation
The recommended printer model is the Zebra 2844 with a serial cable connection. Printouts should be made in accordance to
your facility’s standard operating procedures. Printers are available for purchase through Revolutionary Science. Printer
paper must accommodate printout width of 4 inches. Follow printer manufacturer’s recommendations for regular repair,
cartridge replacement and service.
The printer must be powered and connected to the Saniclave 200 via a serial cable prior to use. Sufficient paper must be
installed in the printer. At the end of the cycle (whether it is a complete cycle or a failure), the SET button, found on the
control panel may be pressed to obtain the printout. If the Saniclave 200 is reset or powered down, the printout data
cannot be recovered.
Sample printout. A successful printout should look similar to the following: