Recommended Load Configuration

Load including the trays must not exceed 4.5 lbs.  

Cycle Phases

Fill

Display reads “Fill” if the 'START' button is pressed and the chamber has
not yet been filled with sufficient water.

Preheat Mode.

Display will read 'Ph' for Preheat.  This mode will usually last for 
about 30 minutes, depending on the load size.

Sterilization Mode

Display will show temperature in Celsius: 

Press ADVANCE to view the pressure reading in PSIg.

Press SET to view the cycle time remaining.

Dry Mode

When the sterilization cycle is successfully completed, the display 
will read 'dry,' for drain mode.  As the water drains, the chamber 
completely depressurizes and shortly thereafter (if selected), the 
Dry Mode begins.  Display will alternate between “dry” and the 30 
minute timer.  When the display reads “turn door,” the door must 
be opened by rotating the door counter clockwise to the “open/dry” 
position until the Dry Mode is complete.

CC Mode

After a successful completion of the entire cycle, the display will 
read 'CC' for Cycle Complete.  Remove the door and remove the 
instruments.  Instruments may be warm to the touch.  

Print

Date/time, preset and achieved parameters are printed only if the 
printer is on and properly connected to the printer port.

8

Содержание RS-SC-200

Страница 1: ...The Saniclave 200 Instruction Manual RS SC 200...

Страница 2: ...10 Warranty Registration and Policy 12 Specifications 13 Contact Information 13 Checklist and Notes of Service Record 14 Care Storage and Maintenance 15 Sterilization Monitoring Guidelines 16 NOTE Re...

Страница 3: ...clave 200 will provide years of reliable service The high quality of construction and durable parts deliver a performance equivalent to much costlier equipment You are invited to visit our web site at...

Страница 4: ...3...

Страница 5: ...the checklist see instruction manual If the shipment is incomplete please contact us at support revsci com or 651 257 0633 3 Carefully read this instruction manual before operating the Saniclave 200...

Страница 6: ...WAYS use ONLY distilled water and never fill chamber with more or less than the recommended amount of distilled water 13 NEVER reuse water left in the exhaust tank Re used water may contain endotoxins...

Страница 7: ...ack Never exceed the maximum load Exceeding maximum load may result in unsterile equipment 8 Run a biological indicator spore test Geobacillus stearothermophilus at least weekly to confirm proper ster...

Страница 8: ...e preheat mode about 25 minutes 13 Observe the display If an error warning is displayed the cycle will be aborted see troubleshooting section on page 10 14 Sterilization Mode After the preheat mode th...

Страница 9: ...tion cycle is successfully completed the display will read dry for drain mode As the water drains the chamber completely depressurizes and shortly thereafter if selected the Dry Mode begins Display wi...

Страница 10: ...ould be used for wrapped loads Note For wrapped loads the user must use wraps or pouches that are cleared by the FDA for human use Unwrapped loads must be used immediately upon sterilization Set Time...

Страница 11: ...Alarm If temperature exceeds the maximum of 124 C the display will read Err 3 Allow unit to drain the reservoir fully into the exhaust tank or sink Press SET to acknowledge the error Open door when c...

Страница 12: ...r the following Power Outage If a power outage occurs the display will revert to standby mode and read 0000 It will not display CC until the Saniclave 200 is restarted and a successful cycle is passed...

Страница 13: ...______________________________________________________________________ Where did you hear of our product ____________________________________________________________________________ Comments _________...

Страница 14: ...ng temperature 121 124 C Temperature lower tolerance 121 C Temperature upper tolerance 124 C Preheat time Automatic Sterilization time 30 minutes Dry time 30 minutes Capacity 12 5 Liters Further Quest...

Страница 15: ..._________________________________________ Your Notes ____________________________________________________________________________ _____________________________________________________________________...

Страница 16: ...e 200 1 Disconnect Saniclave 200 from mains and printer 2 Drain all water from the chamber and exhaust water tank To drain the water exhaust tank turn the manual drain valve located in the rear right...

Страница 17: ...terile and the load configuration and sterilizer shall be investigated due to the fact that proper cycle parameters may have not been correctly achieved Note Improper wrapping packing and loading may...

Страница 18: ......

Отзывы: