Operation and Maintenance Manual – Steam sterilizers mod. BABY S
Rev. 01 del 01-07-2013 - C:\Documenti\Manuali di istruzione\BABY S GB-R01.doc
Pag. 15 di 28
7.1 Sterilizing table
CYCLES FREE
STERILIZING TIME (Min.)
DRYING TIME (Min.)
1
*
*
2
*
*
CYCLES AT 121°
STERILIZING TIME (Min.)
DRYING TIME (Min.)
3 UNWRAPPED
15
5
4 WRAPPED
20
20
CYCLES AT 134°
STERILIZING TIME (Min.)
DRYING TIME (Min.)
5 UNWRAPPED
5
5
6 WRAPPED
10
20
7 PRIONE
18
20
TEST
STERILIZING TIME (Min.)
DRYING TIME (Min.)
BOWIE & DICK*
3,5
2
CYCLE PHIAL
STERILIZING TIME (Min.)
54°
48h (Total length of test)
* Preset value.
7.2 Notes and warnings
1- Strictly comply with the manufacturer's instructions when sterilizing materials.
2- To preserve packaged and loose sterilized materials, the trays must be stored in special cabinets fitted
with ultraviolet lamps and designed for this specific use.
3- Whenever a sterilizing process is carried out, it is advisable to add a chemical indicator that we can
supply on request, code 1468.
4- It is advisable to purchase a printer with date stamp code M 9003 to file and record the sterilizing
cycles and any faults that may have occurred.
5 - There must always be trays and tray-holders in the chamber. Without these accessories, abnormal
pressures and temperatures could develop inside and these could cause the cycle to abort.
6 - It is advisable to place a phial containing culture spores in the chamber along with each load sterilized
in order to make sure that the sterilizing process is valid.
7 - If the packets are not perfectly dry at the end of the sterilizing process, proceed with the supplementary
drying cycle with cycle (DRY FN).