Operation and Maintenance Manual – Steam sterilizers mod. BABY S
Rev. 01 del 01-07-2013 - C:\Documenti\Manuali di istruzione\BABY S GB-R01.doc
Pag. 12 di 28
READ OF INFO DISPLAY:
3
1 5
5
3
= Cycle number
15
= Sterilization time
5
= Dry time
121
= Temperature of sterilization °C
..
1 2 1
4
2 0
1 5
4
= Cycle number
20
= Sterilization time
15
= Dry time
121
= Temperature of sterilization °C
..
1 2 1
5
4
5
5
= Cycle number
4
= Sterilization time
5
= Dry time
134
= Temperature of sterilization °C
..
1 3 4
6
1 0
1 5
6
= Cycle number
10
= Sterilization time
15
= Dry time
134
= Temperature of sterilization °C
..
1 3 4
7
1 5
2 0
7
= Cycle number
15
= Sterilization time
5
= Dry time
134
= Temperature of sterilization °C
..
1 3 4
B & D
T E S T
BOWIE & DICK TEST
not for instruments
134
= Temperature of sterilization °C
..
D R Y
F N
ADDITIONAL DRY
110
= Temperature of sterilization °C
..
B O T
T L E
PHIAL CYCLE
54
= Temperature of sterilization °C for 48 hour
..
N I
G H T
F N
NIGHT CYCLE
YES
= Active
NO
= Inactive
Y E S
/
N O
P
R I
N T E
R
PRINTER SELECTION
YES
= Active
NO
= Inactive
Y E S /
N O
L A N G U A G E
LANGUAGE SELECTION
..
N .
C Y C L E S
NUMBER OF CYCLES
..
W A I
T
>
6
5
Chamber very hot. To attend the cooling.
P
L E A S
E