Packaging, Transport and Installation
19
4.3 Temperature Fluctuations and Condensation
NOTICE
N7.0016
Damaged components due to condensation
Temperature fluctuations
−
The device may be exposed to substantial fluctuations in temperature
during transport. The ensuing condensation can damage electronic
components.
•
Wait until the device has acclimatised before putting it into service.
Temporary storage:
Also in case of an interim storage the device must be stored dry and within the specified
ambient temperature range.
4.4 Conditions for the Installation Site
−
Installation height: max. 2
000 m above sea level
−
Ambient temperature: 5 °C
– 40 °C
NOTICE
N8.0021
Ambient temperature
Temperatures outside the permitted range
−
Electronic and mechanical components may be damaged.
−
The performance data alter to an unknown extent.
•
Do not exceed or fall below the permitted temperature range (5 °C to
40 °C ambient temperature) of the device.
−
Maximum relative humidity < 80 % (at ambient temperatures
≤ 31 °C)
For ambient temperatures U
T
between 31 °C and 40 °C, the maximum relative humidity value L
F
linearly decreases according to L
F
=
–(U
T
– 55) / 0.3:
Ambient temperature
Max. rel. humidity
≤ 31 °C
80 %
33 °C
73.3 %
35 °C
66.7 %
37 °C
60 %
39 °C
53.3 %
40 °C
50 %