Error Messages and Information Notes
43
Pos : 23.1 /0000 Firmenneutral/ 0030 Ü berschrif ten/ 1. Ü berschrif ten/ 1 F ehlermel dungen und Hinw eise @ 13\m od_1448458910002_9.docx @ 93441 @ 1 @ 1
11 Error Messages and Information Notes
Pos : 23.2 /0005 RET SCH/ 0030 Ü bers chrift en/ 1. 1 Ü bersc hrift en/11 F ehlerm eldung en @ 0\m od_1222344568006_9. doc x @ 2435 @ 2 @ 1
11.1 Error Messages
Pos : 23.3 /0000 Firmenneutral/ 0000 M odul e/ 8001 Erkl ärung Fehl ermeldungen @ 16\m od_1465368566287_9. doc x @ 107638 @ @ 1
Error messages inform the user about detected device or programme errors. In the event of an
error message, a fault has occurred, in which the operation of the device or the programme is
automatically interrupted. Such faults must be resolved before next startup.
Pos : 23.4 /0005 RET SCH/ 0005 Bedienungs anl eit ung en/AS 200 / AS 300 seri e/Module/8000 F ehlerm eldung en AS 200 digit c A/c ontrol / AS 300 contr ol B @ 13\m od_1449153595630_9.docx @ 94788 @ @ 1
Error code
Description
Measures
E00
No errors in memory
The test of the memory has been completed
without error.
E10
Drive overload
Switch off the main switch and wait for 30 s
before switching on again.
If the error persists, contact service.
E26
Failure frequency
converter
Switch off the main switch and wait for 30 s
before switching on again.
If the error persists, contact service.
E45
Failure acceleration
sensor
Switch off the main switch and wait for 30 s
before switching on again.
If the error persists, contact service.
Pos : 23.5 /0005 RET SCH/ 0030 Ü bers chrift en/ 1. 1 Ü bersc hrift en/11 Hi nw eise @ 13\mod_1448459519628_9.docx @ 93453 @ 2 @ 1
11.2 Information Notes
Pos : 23.6 /0000 Firmenneutral/ 0000 M odul e/ 8006 Erkl ärung Hinw eis e @ 16\m od_1465368649199_9. doc x @ 107650 @ @ 1
Notices inform the user on specific device or programme processes. The operation of the
device or programme may be interrupted briefly, but there is no fault. The information notice
must be acknowledged by the user to continue the process. Information notices provide
additional information for the user as an aid, but do not represent any device or programme
errors.
Pos : 23.7 /0005 RET SCH/ 0005 Bedienungs anl eit ung en/AS 200 / AS 300 seri e/Module/8005 Hi nw eise AS 200 / AS 300 contr ol B @ 13\m od_1449153613679_9.docx @ 94801 @ @ 1
Notice code
Description
Measures
bS
Display of the complete
runtime in hhh
Press the
button to exit the display.
CAL
Calibration required
Press the
button to exit the display.
Contact the service to arrange an
appointment for the calibration.
dA
Current date required
The device was not connected to the mains or
off for more than 30 days, whereby the battery
is discharged and the current date has been
deleted.
Press the
button. The time display
shows the month "01". The amplitude
display shows the year "-00".
Use the respective "+" and "-" buttons to set
the current month (e.g. "04" for April) and
the current year (e.g. "-16" for 2016).
Press the
button to save the current
date and to exit the display.
Leave the device connected to the mains
and powered on for at least two hours, in
order to fully charge the battery.