Technical Data
15
Pos : 7. 22 /0005 RET SCH/ 0030 Ü bers chrift en/ 1. 1 Ü bersc hrift en/11 Aufgabegröße @ 3\m od_1302161462732_9. doc x @ 24620 @ 2 @ 1
4.8 Feed Grain Size
Pos : 7. 23 /0005 RET SCH/ 0000 Ger ät eneutral/M odul e/ 1040 Aufg abegröß e Analys ensi ebe @ 26\m od_1505395904484_9.docx @ 220396 @ @ 1
Traditional dry sieving is performed in the particle size range of 40 µm to 125 mm. By means of
sieving aids or with wet sieving the measurement range can be extended to 20 µm. The
maximum feed grain size depends on the sample material, the number and aperture size of the
test sieves and the type of the sieving machine.
Examples for the maximum feed grain size according to DIN 66165 are listed in the following
table:
Mesh size
Max. feed grain size
according to DIN 66165
Mesh size
Max. feed grain size
according to DIN 66165
22 µm
710 µm
4 mm
25 mm
45 µm
1 mm
8 mm
45 mm
63 µm
1.4 mm
16 mm
71 mm
125 µm
2.5 mm
22.4 mm
90 mm
250 µm
4 mm
45 mm
150 mm
500 µm
6 mm
63 mm
180 mm
1 mm
10 mm
90 mm
230 mm
2 mm
16 mm
125 mm
300 mm
Pos : 7. 24 /0005 RET SCH/ 0005 Bedienungs anl eit ung en/AS 200 / AS 300 seri e/Module/1041 Aufgabegröße AS 300 c ontrol B @ 16\mod_1461155625919_9. docx @ 105359 @ @ 1
The Vibratory Sieve Shaker AS 300 control is designed for the measurement range of 20 µm to
40 mm.
Pos : 7. 25 /0005 RET SCH/ 0030 Ü bers chrift en/ 1. 1 Ü bersc hrift en/11 Z ul adung @ 14 \mod_1454659163441_9. docx @ 98178 @ 2 @ 1
4.9 Payload
Pos : 7. 26 /0005 RET SCH/ 0005 Bedienungs anl eit ung en/AS 200 / AS 300 seri e/Module/1045 Z ul adung AS 300 c ontrol B @ 50 \mod_1596538868048_9. docx @ 379890 @ @ 1
Voltage
100 V
200 V
230 V
Maximum sample quantity:
3 kg
4 kg
5 kg
Maximum sieve stack weight: 7 kg
9,5 kg
11 kg
Maximum payload:
10 kg
13,5 kg
16 kg
The maximum payload is calculated from the amount of sample quantity (sample material) and
sieve stack mass (test sieves and clamping device) depending on the available nominal voltage.
For the calculation, a "comfort" sieve tensioning unit was used, the weight of which is approx. 5
kg.
−
Maximum sieve stack height:
510 mm
−
Maximum number of fractions: 10 (height of test sieves and collecting pan: 40 mm) /
10 (height of test sieves and collecting pan: 50 mm (2")) /
19 (height of test sieves and collecting pan: 25 mm (1"))
Pos : 7. 27 /0005 RET SCH/ 0030 Ü bers chrift en/ 1. 1 Ü bersc hrift en/11 Verwendbare Siebdurc hmesser @ 14\mod_1454662558474_9. docx @ 98228 @ 2 @ 1
4.10 Suitable Sieve Diameters
Pos : 7. 28 /0005 RET SCH/ 0005 Bedienungs anl eit ung en/AS 200 / AS 300 seri e/Module/1050 Verwendbare Siebdurc hmesser AS 300 c ontrol B @ 16\mod_1461141767951_9. docx @ 105264 @ @ 1
−
Suitable sieve diameters: 100 mm / 150 mm / 200 mm / 203 mm (8") / 305 mm (12") /
315 mm
Pos : 8 / 0000 Firmenneutr al/0000 Module/------- Seit enum bruch ------- @ 9\m od_1427185989165_0.docx @ 76350 @ @ 1
Содержание AS300
Страница 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...
Страница 5: ...Notes on the manual 5 ...
Страница 56: ......
Страница 57: ...Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany ...