Controlling the Device
34
The programming can be cancelled by pressing the
button. All settings are discarded.
7.6.3 Save a Programme
Press the
button (
H8.2
) to save the set sieving parameters in the selected programme
memory slot. The displays stop flashing.
7.7 Signal Tone
The end of the sieving process is announced with an acoustic signal.
Simultaneously press the
(
H6
) and
(
H1
) button to turn off the signal tone. The
process is confirmed by one single signal tone.
Simultaneously press the
(
H6
) and
(
H2
) button to turn on the signal tone. The
process is confirmed by two single signal tones.
7.7.1 Operating Hours
Simultaneously press the
button (
H3
) and the "-" button of the time display (
H5
).
The time display (
H5
) indicates "bS" (Betriebsstunden = operating hours) and the amplitude
display (
H4
) displays the complete runtime (corresponds to the accumulated sieving duration) of
the device in hhh format. All buttons, except for the
button (
H1
) are now locked.
Press the
button (
H1
) to exit the display of the operating hours.
7.7.2 Software Version
Simultaneously press the
button (
H3
) and the "+" button of the time display (
H5
).
The time display (
H5
) indicates "S" (software) and the amplitude display (
H4
) displays the
current number of the software version. All buttons, except for the
button (
H1
) are now
locked.
Press the
button (
H1
) to exit the display of the software version.
7.8 Date
By default, the reminder date for the next calibration is set to the last calibration date plus one
year. However, the reminder date can be changed arbitrarily by the user.
Hold the
button (
H6
) for 5 seconds to display the current reminder date. The current
reminder month is displayed in the time display (
H5
), whereas the current reminder year is
displayed in the amplitude display (
H4
).
Press the "+" or "-" button of the time display (
H5
) to set the desired reminder month.
Press the "+" or "-" button of the amplitude display (
H6
) to set the desired reminder year.
Press the
button (
H1
) to save the settings and to exit the display.
Once the reminder date is reached, the information note "CAL" appears for approx. three
seconds in the amplitude display (
H4
) each time the device is switched on.
Содержание AS 400 control
Страница 2: ...2 Copyright Copyright by Retsch GmbH Retsch Allee 1 5 42781 Haan Germany...
Страница 5: ...Notes on the manual 5...
Страница 49: ......