iBreeze BPAP Systems User Manual
22
Alert
Type
Possible Cause
Action
Water level is too low,
please add to the
appropriate water
level.
Alert 0
During the heating
process, the water level
is below the defined
threshold.
Remove the water
tank and add to the
appropriate water
level.
System error code:
XXXX Please try to
restart, please
contact the supplier if
repeat.
Alert 2
1. Pressure sensor
failure, flow sensor
failure, blower failure in
therapy state; 2. Power
board short-circuit 3.
Humidifier heating
circuit short-circuit
Please try to restart,
please contact the
supplier if repeat.
The respiration tube
expired, please
replace.
Notification In standby state, the
usage time exceed the
respiration tube setting
service life.
Click “Confirm”,
replace the
respiration tube.
The water tank
expired, please
replace.
Notification In standby state, the
usage time exceed the
water tank setting
service life.
Click “Confirm”,
replace the water
tank.
The filter expired,
please replace .
Notification In standby state, the
usage time exceed the
filter setting service life.
Click “Confirm”,
replace the filter
The mask expired,
please replace.
Notification
In standby state, the
usage time exceed the
mask setting service life.
Click “Confirm”,
replace the mask
It is time for device
maintenance ,Please
contact the service
provider for device
maintenance.
Notification
In standby state, the
usage time exceed the
setting device
maintenance time.
Click “Confirm”,
contact the service
provider for device
maintenance.
Low Minute Volume
(MV).
Alert 0
Minute volume less than
the setting threshold
value.
Check the respiration
tube or adjust
therapy parameter
setting.
High Respiratory
Rate (RR).
Alert 0
Respiratory Rate
exceed the setting
threshold value.
Check the respiration
tube or adjust
therapy parameter
setting.
Low Respiratory Rate
(RR).
Alert 0
None.
Check the respiration
tube or adjust
therapy parameter