iBreeze BPAP Systems User Manual
16
Icon
Text
Description
Preheat
Turn Preheat function on or off. When the Preheat is on,
the humidifier starts preheating in the standby mode and
the maximum preheating time is 30 minutes. In the
therapy mode, the preheating stops.
Option: On / OFF Default: OFF
Note:
1. If the humidifier water level below the limit, Preheat
function will be turned off automatically.
2. If humidifier function is turned off, Preheat function will
be disable.
P Ramp
P Ramp: Ramp Starting pressure.
Set the Starting pressure of Ramp function
Setting Range: 3 cmH
2
O-Setting pressure, 0.5 cmH
2
O
increments.
Default: 4 cmH
2
O
Auto Ramp
Turn Auto Ramp function on or off.
When Auto Ramp is on, Sleep state can be identified
automatically, sleep begins to start Ramp function, awake
do not start ramp function.
Option: On / OFF Default: OFF
Auto Start
Turn Auto Start function on or off.
When the Auto Start function is on, the system will start
therapy automatically if a breath with mask is detected.
During therapy, if the mask is removed more than 5
seconds, the therapy mode will stop.
Option: On / OFF Default: OFF
IPR
Setting IPR (Intelligence Pressure Release) level.
Option: OFF / 1-3 Default: 2
Ramp Time
Set the increase time from Ramp pressure to the setting
therapy pressure.
Option: 0-60 mins, 5 mins increments
Default: 15 mins
Note: if Ramp Time sets to 0 min, Ramp function will be
off.
Tube Type
Set the Tube type.
Option: 15 mm / 19 mm Default: 19 mm