iBreeze CPAP Clinical Guide
14
7 Therapy Information
To set the therapy mode, click the Therapy Mode button on the clinical standby interface to enter
the therapy mode menu.
Therapy Mode Setting Interface
Uni-Level therapy mode
In Uni-Level therapy mode, the treatment pressures in inspiration and expiration are equal. The
following modes are Uni-level therapy mode: CPAP, CPAP-V, APAP and APAPW.
7.1 CPAP Mode
In CPAP mode, a constant pressure is delivered to the patient throughout the whole therapy.
When E-COMP is enabled, the treatment pressure delivered at the first day is reduced to the
maximum value of 50% of the prescribed pressure and 4 cmH2O, then treatment pressure
increases 1 cmH2O per day, until the prescribed pressure is reached.
CPAP Mode Parameters Setting
#
Parameter Description
1
Pressure
Set the prescribed pressure throughout the therapy. You can adjust
this setting from 4 to 20 cmH2O. The increment is 0.1/0.5 cmH2O.
2
E-COMP
This feature reduces the therapy pressure setting for the first few days
of operation and gradually increases this setting until the prescription
therapy pressure is reached.
OFF: E-COMP is deactivated.
ON: E-COMP is activated. The initial treatment pressure delivered at
the first day is reduced to the maximum value of 50% of the
prescription pressure, but no lower than 4 cmH2O. After each day of
successful use, the therapy pressure will increase by 1 cmH2O until
the prescription pressure is reached. From that point forward, the
therapy device would operate in normal CPAP mode.