803350 Rev. B                                                                 

 

                                  October 2020 

 
 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Respond O

2

 

 
 

Ultrasonic Oxygen Indicator 

Operation Manual 

 
 
 

Responsive Respiratory, Inc. 

6244 Lemay Ferry Rd. 

St. Louis, MO 63129 

 

[email protected]

 

www.respondo2.com

 

866-333-4030 

 
 
 
 
 

 

Содержание Respond O2

Страница 1: ...e 866 333 4030 Fax 866 333 4035 sales respondo2 com www respondo2 com 803350 Rev B October 2020 Respond O2 Ultrasonic Oxygen Indicator Operation Manual Responsive Respiratory Inc 6244 Lemay Ferry Rd S...

Страница 2: ...turn the unit to Responsive Respiratory Responsive Respiratory will be responsible for normal shipping charges for the return of the unit to purchaser All warranties are voided if the Respond O2 unit...

Страница 3: ...ration from any source other than a conventional oxygen concentrator using molecular sieve beds This unit will not withstand fluid spillage The unit will not withstand mechanical shock or vibration NO...

Страница 4: ...ies Clean battery terminals When unit is turned on display flashes PRO O2 CAL and then turns off Unit requires calibration Perform Full Calibration Procedure outlined in Operation Manual Display reads...

Страница 5: ...of an oxygen concentrator adjusting for the argon that passes through the sieve beds 2 When using the unit with the pure O2 calibrate button depressed the software measures oxygen not adjusting for th...

Страница 6: ...s Will the unit lose calibration when I change batteries No The unit holds calibration even when the battery is removed 12 Ultrasonic Oxygen Indicator FREQUENTLY ASKED QUESTIONS The following are some...

Страница 7: ...t supply tubing to the concentrator oxygen outlet and the inlet port 5 Turn on the device See above 6 The indicator will stabilize in about 10 seconds and provide an accurate reading on the LCD displa...

Страница 8: ...ircuit will compensate for variations in temperature The temperature during storage and shipping should not reach below 29 F 34 C or above 160 F 71 C If the unit has been stored at other than normal r...

Страница 9: ...cator FULL TWO STEP CALIBRATION PROCEDURE The following procedure may be used to calibrate the indicator to both room air and U S P oxygen 99 0 1 Do not connect unit to oxygen source or introduce oxyg...

Страница 10: ...the rear of the unit away from the case Connect the 9 volt battery to the standard 9 volt battery connection and lay the battery sideways in the battery compartment Insert battery door by sliding door...