9
5.
Zip case fully closed (or secure lid, if utilizing container). The sanitizing
cycle for Purify O3 Elite is approximately 35 minutes in length. During the
cycle, the Progress Indicator Ring will darken to indicate when the cycle
is one fourth, halfway, and three-fourths complete as shown in Figure 10.
At the conclusion of disinfection, the sanitizer will beep once and power
off.
The sanitizer and accessories must remain in the closed case/container
for an additional 90 minutes to allow the residual ozone to dissipate.
Before use, run your PAP machine for 20-60 seconds with mask and
tubing attached to flush fresh air through accessories.
Your CPAP/BiPAP humidification chamber may be sanitized weekly, as
desired.
NOTE:
To stop the sanitation process mid-cycle, place finger on the On/Off
Touch Sensor for 4 seconds. The sanitizer will signal the cancellation with
a long beep and power off. Reseal case/container and let sanitizer and
accessories sit in closed case/container for at least 90 minutes to allow
residual ozone to dissipate before use. Stopping a sanitizing cycle before
completion is not recommended.
IMPORTANT:
Do not use your CPAP/BiPAP accessories for at least
90 minutes after sanitation is complete to allow
remaining ozone to dissipate.
WARNING:
Operate only in a well ventilated area.
Do not breathe or ingest the ozone.
Figure 10
1/4 cycle
1/2 cycle
3/4 cycle