9

5. 

Zip case fully closed (or secure lid, if utilizing container). The sanitizing 

cycle for Purify O3 Elite is approximately 35 minutes in length. During the 

cycle, the Progress Indicator Ring will darken to indicate when the cycle 

is one fourth, halfway, and three-fourths complete as shown in Figure 10.  

At the conclusion of disinfection, the sanitizer will beep once and power 

off. 

 

 

 

The sanitizer and accessories must remain in the closed case/container 

for an additional 90 minutes to allow the residual ozone to dissipate. 

Before use, run your PAP machine for 20-60 seconds with mask and 

tubing attached to flush fresh air through accessories.

 

Your CPAP/BiPAP humidification chamber may be sanitized weekly,  as 

desired.

NOTE:  

To stop the sanitation process mid-cycle, place finger on the On/Off 

Touch Sensor for 4 seconds. The sanitizer will signal the cancellation with 

a long beep and power off. Reseal case/container and let sanitizer and 

accessories sit in closed case/container for at least 90 minutes to allow 

residual ozone to dissipate before use. Stopping a sanitizing cycle before 

completion is not recommended. 

IMPORTANT:  

Do not use your CPAP/BiPAP accessories for at least 

90 minutes after sanitation is complete to allow 

remaining ozone to dissipate.

WARNING:  

Operate only in a well ventilated area. 

Do not breathe or ingest the ozone.

Figure 10

1/4 cycle

1/2 cycle

3/4 cycle

Содержание 190-6100

Страница 1: ...866 333 4035 F www purifyo3 com www respondo2 com Warnings Specifications 1 Parts Components 2 Prior to Use Important Notes Warranty 3 Charging the Battery 4 Mask Tubing Tees Parts Sanitation Deodorizing 5 7 CPAP BiPAP Humidification Chamber Sanitation 8 9 Frequently Asked Questions 10 Para instrucciones en español refiérase a las páginas 11 21 ...

Страница 2: ...tructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures R...

Страница 3: ... closed undisturbed for at least 90 minutes after the sanitation cycle is complete to allow for the dissipation of remaining ozone Do not use equipment for at least 90 minutes after sanitation cycle is finished SYMBOLS This device complies with part 15 of FCC rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept an...

Страница 4: ...g Adapter 190 6004 USB Cable 190 6103 Dual Purpose Sanitizing Travel Case 190 6002 Tubing Adapter 190 6005 Purify O3 Elite Sanitizer Replacement component parts available for purchase separately contact your local retailer for availability and pricing USB Charging Port Tubing Port On Off Touch Sensor Battery Life Indicator Disinfection Progress Ring SANITIZER COMPONENTS ...

Страница 5: ...sanitizing to allow the residual ozone to dissipate back to oxygen Ensure the Purify O3 Elite is operated in a well ventilated area Do not breathe or ingest the ozone release during after operation Warranty is VOID if repairs modifications to the device are made by unauthorized persons To obtain service within the warranty period please contact the retailer where unit was purchased Purify O3 Elite...

Страница 6: ... with a series of blinking white indicator lights When the Purify O3 Elite is fully charged the Battery Life Indicator lights will change to solid white Charging time averages 2 3 hours and may vary depending on the battery s remaining capacity If the sanitizer stops working during use after a full charge the battery may have reached the end of its serviceable life LOW POWER WARNING When low batte...

Страница 7: ...t CPAP tubing directly to the Purify O3 Elite as shown in Figure 2 If the accessories are too large for the sanitizing case a medium sized storage container with tightly fitting lid 12 qts or less may be used refer to Figure 3 below NOTE Purify O3 Elite is intended for use to sanitize and deodorize accessory items only Please refer to your user manual for guidance on cleaning and disinfection of t...

Страница 8: ...cycle for Purify O3 Elite is approximately 35 minutes in length During the cycle the Progress Indicator Ring will darken to indicate when the cycle is one fourth halfway and three fourths complete as shown in Figure 6 At the conclusion of disinfection the sanitizer will beep once and power off The sanitizer and accessories must remain in the closed case container for an additional 90 minutes to al...

Страница 9: ...r off Reseal case container and let sanitizer and accessories sit in closed case container for at least 90 minutes to allow residual ozone to dissipate before use Stopping a sanitizing cycle before completion is not recommended IMPORTANT Do not use your mask tubing or accessories for at least 90 minutes after sanitation is complete to allow remaining ozone to dissipate ...

Страница 10: ...ease refer to your CPAP BiPAP machine s user manual for guidance on cleaning and disinfection of the PAP machine 4 To activate the Purify O3 Elite place and hold finger on the On Off Touch Sensor for 2 seconds as shown in Figure 8 The Purify O3 Elite will beep once and the Disinfection Progress Ring shown in Figure 9 will illuminate to indicate the disinfection process has begun The Battery Life I...

Страница 11: ...ing attached to flush fresh air through accessories Your CPAP BiPAP humidification chamber may be sanitized weekly as desired NOTE To stop the sanitation process mid cycle place finger on the On Off Touch Sensor for 4 seconds The sanitizer will signal the cancellation with a long beep and power off Reseal case container and let sanitizer and accessories sit in closed case container for at least 90...

Страница 12: ...nitizer is less than 2 years from date of purchase contact the retailer the sanitizer was purchased from for warranty evaluation If over 2 years from purchase dispose of the product per electronic recycling guidelines in your area Purify O3 Elite does not power off automatically after the sanitation cycle Power the sanitizer off by covering the On Off Touch Sensor for 4 seconds If the problem pers...

Страница 13: ...do de agua Este equipo tiene componentes electrónicos Uselo solamente en área bien ventilada No lo use cerca de llamas vapor de gas u otros combustibles Manténgalo fuera del alcance de los niños No respire o ingiera ozono liberado por el equipo Permita que la estuche o contenedor de desinfección permanezca cerrado sin molestias durante al menos 90 minutos después de que se complete el ciclo de des...

Страница 14: ...separado contacte a su vendedor minorista local para disponibilidad y precios Entrada de carga USB Entrada del tubo El sensor táctil de encendido apagado Indicador de carga Anillo de progreso de desinfección COMPONENTES DE LA UNIDAD PARTES INCLUIDAS Purify O3 Elite unidad desinfectante con batería recargable Cable USB y adaptador de carga Adaptador de tubo Estuche de doble uso disenfector viaje Ma...

Страница 15: ... accesorios hasta por lo menos depues de pasados 90 minutos de desinfección para permitir que el ozono producido se dicipe de nuevo al oxígeno Asegúrese que el Purify O3 sea operado en un área bien ventilada No respire o ingiera el ozono disipado durante después de la operación La garantía será ANULADA si reparaciones modificaciones han sido realizadas por personas no autorizadas Para obetener ser...

Страница 16: ...icadoras blancas parpadeantes Cuando el Purify O3 Elite está completamente cargado las luces indicadoras de duración de la batería cambiarán a blanco sólido El tiempo promedio de carga es de 2 3 horas y puede variar dependiendo en la capacidad restan te en la batería Si la unidad deja de funcionar mientras está en uso depués de haber cargado com pletamente la batería puede haber llegado al fin de ...

Страница 17: ...ra 2 Si el conjunto del tubo y la máscara es demasiado grande para la caja de desinfección un contenedor de almacenamiento tamaño mediano con tapa sólida 12 cuartos o menos puede usarse consulte la Figura 3 a continuación Para el saneamiento solo con máscara coloque la máscara y Purify O3 Elite dentro de la caja de desinfección almacenamiento NOTA Purify O3 Elite está diseñado para desinfectar los...

Страница 18: ...ciclo de desinfección para Purify O3 Elite es de aproximadamente 35 minutos de duración Durante el ciclo el Anillo indicador de progreso se oscurecerá para indicar cuándo el ciclo está completo en un cuarto a la mitad y en tres cuartos como se muestra en la Figura 6 Al finalizar la desinfección el desinfectante emitirá un sonido y se apagará El desinfectante y los accesorios deben permanecer en la...

Страница 19: ...l ciclo coloque el dedo en el sensor táctil de encendido apagado durante 4 segundos El desinfectante indicará la cancelación con un pitido largo y apagado Vuelva a sellar la caja recipiente y deje que el desinfectante y los accesorios se asienten en la caja recipiente cerrado durante al menos 90 minutos para permitir que el ozono residual se disipe antes de usar No se recomienda detener un ciclo d...

Страница 20: ...onsulte el manual de usuario de su máquina CPAP BiPAP para obtener orientación sobre la limpieza y desinfección de la máquina PAP 4 Para activar el Purify O3 Elite coloque y mantenga el dedo en el sensor táctil de encendido apagado durante 2 segundos como se muestra en la Figura 8 El Purify O3 Elite emitirá un pitido una vez y el anillo de progreso de desinfección que se muestra en la Figura 9 se ...

Страница 21: ...sco a través de los accesorios Su cámara de humidificación CPAP BiPAP puede desinfectarse semanalmente según lo desee NOTA Para detener el proceso de saneamiento a mitad del ciclo coloque el dedo en el sensor táctil de encendido apagado durante 4 segundos El desinfectante indicará la cancelación con un pitido largo y apagado Vuelva a sellar la caja recipiente y deje que el desinfectante y los acce...

Страница 22: ...da Conecte el cargador a la unidad y enchúfelo salida para cargar Purificar O3 Elite se detiene durante el uso después de una carga completa La batería puede haber llegado al final de su vida útil Cargue la unidad completamente y intente ejecutar un ciclo completo de desinfección Si el problema persiste y la unidad es inferior a 2 años a partir de la fecha de compra comuníquese con el vendedor don...

Страница 23: ... cámara de humidificación de agua en un ciclo para ahorrar tiempo Siga las instrucciones de desinfección que se enumeran en la sección de instrucciones de este folleto Desinfección de máquinas CPAP BiPAP Purify O3 Elite está diseñado para desinfectar los accesorios CPAP BiPAP Consulte el manual de usuario de su máquina CPAP BiPAP para obtener orientación sobre limpieza y desinfección de la máquina...

Страница 24: ...Responsive Respiratory Inc 6244 Lemay Ferry Road St Louis MO 63129 866 333 4030 866 333 4035 F www purifyo3 com www respondo2 com 190 6106 0520 ...

Отзывы: