background image

2091 Exchange Court 
Fairborn, Ohio 45324 
937-426-2703 
www.NexSens.com 

Observator NEP5000 Turbidity Sensor 

Quick Start Guide 

 

 

The Observator ANALITE NEP5000 Turbidity Sensor is compatible with the Nexsens 
X2 data logger using SDI-12 output, and is designed for environmental monitoring 
and process applications. 

 

 

 

 

 
 
Calibrating the NEP5000 

For configuration and calibration, the Observator ANALITE NEP5000 Configuration 
Kit accessory must be used. This kit comes with a dedicated Subconn connection 
cable. 

The Subconn cable to calibration module connection is detailed in the table below.  

 

 

 

 

 

 

 

Subconn Wire Color

 

Wire Outputs

 

Calibration

 

Module Connection

 

Black 

Power GND 

Slot 2 

White 

SDI-12 Data 

N/A 

Red 

10 to 30V DC power 

Slot 1 

Green 

4-20mA loop to power 

GND 

N/A 

Orange 

Voltage Out 

N/A 

Blue 

Calibration 

Communication Data 

Slot 6 

Отзывы: