Monitoring the Baby
45
Monitoring the Baby
Turning the SmartMonitor 2 on
You have properly set up the SmartMonitor 2, understand how the monitor functions and how to respond
to alarms, you are now ready to begin monitoring the baby’s breathing and heart activity according to the
schedule prescribed by the physician.
Push the
POWER
button. The SmartMonitor 2 performs a system check. The lights on the front of the
SmartMonitor 2 and the alarm come on briefly. Within 10 seconds, the green
respiration
, and
heart
lights
begin to blink. If the lights do not blink, check that you have attached the electrode belt properly to the
baby and that lead wires are pushed in, and that cables are connected.
Once the baby is properly connected to the SmartMonitor 2 and the power is on:
The green (battery)
charger
light is on.
The green
power
light is on.
The green
respiration
light and the green
heart
light are blinking.
All other lights should be off.
If the lights do not blink, refer to the steps found in the Setting Up the SmartMonitor 2 section in
this manual and be sure you have followed all instructions.
Turning the Monitor off - Sibling Alarm
The SmartMonitor 2 has a built-in safety feature called a
sibling
alarm.
If the monitor is not turned off in a
specific sequence, the green power light will remain on and the alarm will sound continuously. This safety
feature makes sure the power is not accidentally turned off. To turn the SmartMonitor 2 off:
Press and hold the blue
RESET
button.
Press and release the gray
POWER
button.
Wait 2 seconds, and then release the
RESET
button.
When the SmartMonitor 2 is turned off without pushing the
RESET
button first, the green power light will
remain on and the Sibling Alarm will sound. To silence the Sibling Alarm:
Press and hold the blue
RESET
button.
Press and release the gray
POWER
button.
Wait 2 seconds, and then release the
RESET
button.
To resume monitoring, press the gray
POWER
button.
If there is an internal software error, a special power off procedure is required.
Press and hold the
RESET
button, while still holding down the
RESET
button press and hold the
POWER
button. Hold both buttons down for 5 seconds.
Release
POWER
button; continue to hold the
RESET
button until the monitor turns off.
Содержание SmartMonitor2
Страница 1: ...Professional Operator s Manual ...
Страница 6: ...Table of Contents ...
Страница 12: ...Introduction 6 ...
Страница 16: ...Warnings and Cautions 10 ...
Страница 36: ...Using Auxiliary Equipment 30 ...
Страница 38: ...Dealer Preparation for Home Setup 32 ...
Страница 42: ...Home Setup 36 Baby s right side Baby s left side ...
Страница 50: ...Reducing False Alarms 44 ...
Страница 54: ...Monitoring the Baby 48 ...
Страница 58: ...Transferring SmartMonitor 2 Information 52 ...
Страница 64: ...Transferring SmartMonitor 2 Information 58 ...
Страница 68: ...Caring for the SmartMonitor 2 62 ...
Страница 72: ...Ordering Information 66 ...
Страница 74: ...Appendices 68 ...
Страница 76: ...Appendices 70 ...
Страница 78: ...Specifications 72 ...
Страница 80: ...H577 4000 10 KK 5 3 06 ...