Setting Alarm and Recording Limits
27
After the new parameters have been entered into the SmartMonitor 2, the monitor must be turned off to
exit Menu Mode.
To set the parameters through the Synergy-E computer software, refer to the
Synergy-E Manual
for
instructions on communicating with the SmartMonitor 2.
System Set-up Menu
SmartMonitor 2 also has a System Set-up Menu, which displays the entire menu of parameters. This
includes very seldom-used menu choices. In addition to some menus previously seen, you will have
access to the following menus. To access this menu, do the following:
Press
the
POWER
button ON. After a short delay, the display will read:
INITIALIZING PLEASE WAIT
Then,
MENU MODE? ENTER PROPER KEY SEQUENCE
Press the keys in the following sequence within 10 seconds:
Press
the
t
Up arrow 3 times.
Press
the
ENTER
button 3 times.
Scroll to MENU OPTIONS and press ENTER.
Scroll to SYSTEM MENU and press ENTER.
The following table lists the additional options available from the System menu:
Menu Option
LCD Display
Description of Option or Instructions
Date Format
DATE FORMAT
Permits the selection of two formats for
displaying the DATE:
MONTH/DAY/YEAR or
DAY/MONTH/YEAR
Standard value:
Month/Day/Year
Memory Full Alert
MEMORY FULL ALERT
Two options are available: 80% full or
50% full.
When the memory usage reaches the
selected limit (50% or 80%), the unit will
generate an alarm.
Standard value:
50%
Memory Full Audible
MEMORY FULL AUDIBLE
Used to enable/disable the audible alarm
when the unit reaches the selected
memory almost full limit (50%/80%).
NOTE: This parameter does not affect the
visual indicator; it cannot be disabled.
Values:
ON, OFF
Standard Value:
OFF
Modem Speed -
Monitor
Modem Speed-Monitor
The default setting is 38,400Bps. This
setting should not be change unless
directed to do so by Respironics
Technical Support.
Содержание SmartMonitor2
Страница 1: ...Professional Operator s Manual ...
Страница 6: ...Table of Contents ...
Страница 12: ...Introduction 6 ...
Страница 16: ...Warnings and Cautions 10 ...
Страница 36: ...Using Auxiliary Equipment 30 ...
Страница 38: ...Dealer Preparation for Home Setup 32 ...
Страница 42: ...Home Setup 36 Baby s right side Baby s left side ...
Страница 50: ...Reducing False Alarms 44 ...
Страница 54: ...Monitoring the Baby 48 ...
Страница 58: ...Transferring SmartMonitor 2 Information 52 ...
Страница 64: ...Transferring SmartMonitor 2 Information 58 ...
Страница 68: ...Caring for the SmartMonitor 2 62 ...
Страница 72: ...Ordering Information 66 ...
Страница 74: ...Appendices 68 ...
Страница 76: ...Appendices 70 ...
Страница 78: ...Specifications 72 ...
Страница 80: ...H577 4000 10 KK 5 3 06 ...