Using Auxiliary Equipment
29
Using Auxiliary Equipment
There are 4 auxiliary channels available on the SmartMonitor 2. These channels provide the mechanism
for connecting External Devices to the SmartMonitor 2. All External Devices must interface to the
SmartMonitor 2 via the Auxiliary Interface Box, P/N 4010. The output of these Devices must be within the
range of –1.25 to +1.25 volts.
Setup of auxiliary channels can only be done through the Synergy-E software. Refer to the
Synergy-E manual for more information.
SmartMonitor 2 is programmed with 15 signal definitions to interpret the auxiliary signals it receives. For
each of the 4 auxiliary channels you can choose from the list of 15. These are user configurable with the
Synergy-E software. Refer to the Synergy-E manual for more information on auxiliary signal configuration
for the SmartMonitor 2.
Any of the four signals can be programmed to record at one time. The Auxiliary Interface Box, AIB,
connects to the I/O Connections port.
Connect the AIB to the back of
the SmartMonitor 2.
Connect appropriate interface
cable to the number 1 slot on the
AIB. The other end of the cable
should be connected to the
auxiliary device.
Insert into the I/O
connector on the back
of the SmartMonitor 2
Connect device
interface cable into
the #1 slot on the AIB
and the other end
into the external
device
Before turning on the SmartMonitor 2 insure that the external devices are connected.
For every signal you want to record, you will need to connect an auxiliary cable to the AIB and
remember to turn on the Auxiliary Channels in the SmartMonitor 2 menu. For example if you
wish to record SpO
2
and Pulse you would need 2 auxiliary cables and this will use 2 auxiliary
channels.
Information required to setup or change the “definition” includes the devices voltage and value range.
This information must come from the Manufacturer of the device.
Using the Respironics Model 930 Oximeter
When the SmartMonitor 2 is interfaced to the 930 Oximeter, it can be configured to trigger recordings
when the Oximeter alarms.
Attach the Auxiliary Interface box to the SmartMonitor 2 and the 4304 cable to the 930 Oximeter. Enter
the System Set-up menu mode and go to Ext. Physio Trigger and set to Trigger When High. When the
Oximeter alarms the SmartMonitor 2 will record all programmed channels that are turned on.
Содержание SmartMonitor2
Страница 1: ...Professional Operator s Manual ...
Страница 6: ...Table of Contents ...
Страница 12: ...Introduction 6 ...
Страница 16: ...Warnings and Cautions 10 ...
Страница 36: ...Using Auxiliary Equipment 30 ...
Страница 38: ...Dealer Preparation for Home Setup 32 ...
Страница 42: ...Home Setup 36 Baby s right side Baby s left side ...
Страница 50: ...Reducing False Alarms 44 ...
Страница 54: ...Monitoring the Baby 48 ...
Страница 58: ...Transferring SmartMonitor 2 Information 52 ...
Страница 64: ...Transferring SmartMonitor 2 Information 58 ...
Страница 68: ...Caring for the SmartMonitor 2 62 ...
Страница 72: ...Ordering Information 66 ...
Страница 74: ...Appendices 68 ...
Страница 76: ...Appendices 70 ...
Страница 78: ...Specifications 72 ...
Страница 80: ...H577 4000 10 KK 5 3 06 ...