P
AGE
5-24
1117539, V
ER
. 07
Oximetry: Good
Connection
(icon only)
SpO
2
Status
Displays on the
therapy screen
when the blower is
on and 3 seconds of
good connection is
detected. Appears
at the beginning of
therapy.This screen
will not display
again if the finger
probe is removed
and reapplied
unless therapy is
stopped and
restarted.
n/a
No action
needed.
Pair?: 123456
Yes/No
Status
Prompts to accept or
decline pairing to a
Bluetooth compatible
device. This device
can be identified by
the digits displayed.
n/a
Rotate control
dial to accept
pairing (Yes), or
decline (No),
then press
control dial to
confirm
selection.
SD Card
Removed.
Notification
or Alert 2
Indicates SD card
has been removed
from therapy device
and not reinserted
before the start of
the current therapy
session.
SD card was not
reinserted into
device.
Reinsert SD
card, or click to
clear alert.
Oximetry:
Good Study
(icon only)
SpO
2
Notification
Notifies that user has
a achieved at least 4
hours of therapy and
oximetry use. Appears
at the end of therapy.
n/a
Press Control
Dial to
acknowledge
and clear the
message.
A
LERT
I
CON
T
YPE
D
ESCRIPTION
P
OSSIBLE
C
AUSE
A
CTION
Содержание DreamStation
Страница 1: ...DreamStation Service Technical Reference Manual ...
Страница 5: ...1117539 VER 07 This page intentionally blank ...
Страница 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Страница 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Страница 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Страница 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Страница 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Страница 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...