Device Settings:
Choosing this screen will take you to a sub-menu where you can adjust the way the device displays
information. These settings are described here.
F
IGURE
4-23: S
AMPLE
D
EVICE
S
UB
-M
ENU
Note:
Not all settings shown here will display on the device. The display will vary based on therapy device
model and device settings.
Mask Type
This setting allows you to select the appropriate Mask Type
resistance setting (also known as System One Resistance
Control) for your Philips Respironics mask. This feature allows
the device to adjust the level of pressure compensation to
match your mask. Refer to the packaging of your mask to
identify the resistance setting for your mask.
Note: It is important to use the appropriate “Mask Type”
resistance setting to ensure proper pressure delivery to the
patient.
Mask Type Lock
This enables you to lock the Mask Type resistance setting if
you do not want the patient to change it.
Check Mask Fit
You can enable or disable the check mask fit setting. This
feature allows the patient to check the fit of their mask prior to
starting therapy. This is done by measuring the amount of leak
in the patient circuit.
I
CON
T
EXT
D
ESCRIPTION
Содержание DreamStation
Страница 1: ...DreamStation Service Technical Reference Manual ...
Страница 5: ...1117539 VER 07 This page intentionally blank ...
Страница 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Страница 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Страница 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Страница 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Страница 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Страница 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...