P
AGE
4-18
1117539, V
ER
. 07
4.10.3 A
CCESSING
P
ROVIDER
M
ODE
S
CREENS
Accessing provider mode unlocks settings that cannot be modified by the user. To access provider mode:
1.
Supply power to the device. First, plug the socket end of the AC power cord into the power supply.
Then plug the pronged end of the AC power cord into an electrical outlet that is not controlled by a
wall switch. Finally, plug the power supply cord’s connector into the power inlet on the back of the
device.
2.
Once the device is powered, press and hold both the control dial and the Ramp button on the
device for at least 5 seconds.
3.
You are now in provider mode. You can choose between the following Provider mode screens.
4.10.4 N
AVIGATING
THE
P
ROVIDER
M
ODE
S
CREENS
The following sections will describe the options available from the Provider screens:
Therapy Settings:
Choosing this screen will take you to a sub-menu where you can adjust the device therapy modes and
pressure settings. These settings are described here.
F
IGURE
4-21: S
AMPLE
T
HERAPY
S
UB
-M
ENU
Note:
Not all settings shown here will display on the device. The display will vary based on therapy device
model and device settings.
Содержание DreamStation
Страница 1: ...DreamStation Service Technical Reference Manual ...
Страница 5: ...1117539 VER 07 This page intentionally blank ...
Страница 6: ...PAGE 1 1 1117539 VER 07 2016 Koninklijke Philips N V All rights reserved ...
Страница 11: ...PAGE 1 6 1117539 VER 07 This page intentionally blank ...
Страница 13: ...PAGE 2 2 1117539 VER 07 This page intentionally blank ...
Страница 15: ...PAGE 3 2 1117539 VER 07 This page intentionally blank ...
Страница 134: ...PAGE 8 8 1117539 VER 07 This page intentionally blank ...
Страница 135: ...NO 1117539 VER 07 EMH 06 04 2016 ...