Setting up for invasive use

En

glish

Plug the free end of the power cord into a power outlet. 

Attach the H4i heated humidifier to the front of the Stellar.

If the H4i is not in use, and if applicable, connect the antibacterial filter firmly onto the air 
outlet of the device (see “Attaching an antibacterial filter” on page 10).

Note:

 Other external humidifiers can be used for noninvasive setup. ResMed 

recommends using a breathing system (including antibacterial filter, air tubings, external 
humidifier) with an impedance of maximum 2 cm H

2

O at 30 L/min, 5 cm H

2

O at 60 L/min 

and 16 cm H

2

O at 120 L/min.

4

Connect one end of the air tubing firmly onto the air outlet of the H4i.

5

Connect the mask system to the free end of the air tubing. 

6

Press the power switch on the back to turn on the device.

7

Select the mask type (select 

Setup

 menu, then 

Clinical Settings

, then 

Advanced Settings

).

8

Perform Learn Circuit (see “Setup menu: Options” on page 18).

Attaching the H4i heated humidifier for noninvasive use

Use of the H4i heated humidifier may be recommended by your clinician. For information 
on using the H4i, see the H4i user guide.

WARNING

Always place the H4i on a level surface below the level of the patient to prevent 
the mask and tubing from filling with water.

Make sure that the water chamber is empty and thoroughly dried before 
transporting the humidifier.

The H4i heated humidifier is not intended for mobile use.

Do not overfill the water chamber, as during use this will cause water to spill into 
the air circuit. Check the air circuit regularly for accumulated water. If you observe 
water has spilled into the air circuit, clear the water and check that the water 
chamber is not overfilled. When using the H4i at high pressures (above 25 cm 
H

2

O) the use of a water trap will avoid water spillage back into the air circuit.

For optimal accuracy and synchrony, perform a Learn Circuit with a change of the 
circuit configuration, in particular when adding or removing high impedance 
components (eg, antibacterial filter, external humidifier, water trap, nasal pillow 
type mask or air tubing). See “Setup menu: Options” on page 18.

CAUTION

Check the air circuit for water condensation. Use a water trap or a tubing wrap if  
humidification is causing water condensation within the tube. 

Notes:

A humidifier increases resistance in the air circuit and may affect triggering and cycling, and 
accuracy of display and delivered pressures. Therefore perform the Learn Circuit function 
(see “Setup menu: Options” on page 18
). The device adjusts the airflow resistance.

The heating feature of the H4i is disabled when the device is not mains powered.

Setting up for invasive use

The Stellar can be used invasively only with the ResMed Leak Valve, or using an uncuffed 
or deflated cuff tracheostomy tube with the ResMed Leak Port (24976).

AutoEPAP is contraindicated for invasive use.

WARNING

Make sure that all the air inlets at the rear of the device and under the device and 
vents at the mask or at the ResMed Leak Valve are unobstructed. If you put the 
device on the floor, make sure the area is free from dust and clear of bedding, 
clothes or other objects that could block the air inlets.

When using a Heated Moisture Exchange Filter (HMEF), replace the HMEF 
regularly as specified in the instructions provided with the HMEF.

Содержание stellar series

Страница 1: ...User Guide English User guide Non invasive invasive ventilators...

Страница 2: ...g the pulse oximeter 8 Adding supplemental oxygen 8 Using the FiO2 monitoring sensor 10 Attaching an antibacterial filter 10 Stellar basics 11 About the control panel 11 LCD screen 12 Starting therapy...

Страница 3: ...isinfection 23 Multipatient use 23 Servicing 24 Troubleshooting 24 Alarm troubleshooting 24 Other troubleshooting 27 Technical specifications 30 Guidance and manufacturer s declaration electromagnetic...

Страница 4: ...endure more than brief interruptions in ventilation The Stellar is not a life support ventilator If you have any of the following conditions tell your doctor before using this device pneumothorax or...

Страница 5: ...ece adapter FiO2 monitoring sensor Stellar Mobility bag ResMed Leak Valve Tubing wrap Stellar DC DC converter WARNING The Stellar should only be used with air tubing and accessories recommended by Res...

Страница 6: ...ry the device settings and the patient circuit configuration In case of a mains power disruption the device will operate using the internal battery if there is no external battery connected to the dev...

Страница 7: ...device in the event of accidental liquid spillage onto the device Ensure that the air filter and air filter cover are fitted at all times Make sure that all the air inlets at the rear of the device a...

Страница 8: ...rly for accumulated water If you observe water has spilled into the air circuit clear the water and check that the water chamber is not overfilled When using the H4i at high pressures above 25 cm H2O...

Страница 9: ...et to Trach AutoEPAP has to be turned off 1 Connect the power cord 2 Plug the free end of the power cord into a power outlet 3 Connect the antibacterial filter firmly onto the air outlet of the device...

Страница 10: ...to the patient side of the ResMed Leak Valve 11 Connect the catheter mount The ResMed Leak Valve or the HMEF can be connected to standardised tracheostomy interfaces including connector pieces like ca...

Страница 11: ...d by your clinician Note Up to 30 L min can be added WARNING Oxygen flow must be turned off when the device is not operating so that unused oxygen does not accumulate within the device and create a ri...

Страница 12: ...y locks into place 3 Attach the other end of the oxygen supply tubing to the oxygen supply 4 Start ventilation 5 Turn on the oxygen source and adjust to the desired flow rate To remove supplemental ox...

Страница 13: ...he air outlet of the device 4 Connect one end of the cable to the FiO2 monitoring sensor 5 Connect the other end of the cable to the rear of the device 6 Start calibration see Setup menu Options on pa...

Страница 14: ...e introduced by the filters Stellar basics About the control panel l 1 2 Mains Power LED On when operating on mains power Internal battery LED On when internal battery is in use Flashes when device is...

Страница 15: ...See Working with alarms on page 15 In stand by mode Extended hold for at least three seconds starts the alarm LED and alarm buzzer test Menu keys Press the appropriate menu button Monitoring Setup In...

Страница 16: ...ttery LED is on Note If the charge state of the internal battery is too low or if the battery is empty an alarm occurs See the Alarm troubleshooting section on page 24 for further information Reconnec...

Страница 17: ...move the patient interface and press to stop airflow or if SmartStart Stop is enabled simply remove the patient interface and treatment will stop automatically Notes SmartStop may not work if Full fac...

Страница 18: ...igh From the Setup menu select Alarm Settings After the set value has been confirmed the alarm will sound and the alarm LED lights Alarm settings see Setup menu Alarm Settings Alarm Volume on page 18...

Страница 19: ...g off the ventilation you can use the Ramp Down feature This feature reduces the treatment pressure to Start EPAP PEEP over five minutes After five minutes the therapy stops You can enable the Ramp Do...

Страница 20: ...d by the equivalent buttons on the right of the LCD screen In each menu there are screens that display settings device or therapy information Note iVAPS mode is only available in Stellar 150 MONITORIN...

Страница 21: ...a full list of compatible masks for this device see the Mask Device Compatibility List on www resmed com on the Products page under Service Support If you do not have internet access please contact y...

Страница 22: ...libration The device starts the calibration of the FiO2 monitoring sensor to measure the oxygen concentration of the breathable air 1 Press to start the FiO2 sensor calibration 2 Wait for the device t...

Страница 23: ...the data of the last 365 days Device Information This screen shows the serial number displayed as Device SN device number software version and other component versions Data on this screen may be reque...

Страница 24: ...hecks if there is enough memory capacity on the USB stick and if readable data is available 2 Select one of the available options Read settings The settings will be transfered from the connected ResMe...

Страница 25: ...ly 1 Remove the air tubing from the device and the patient interface 2 Wash the air tubing in warm water using mild detergent 3 Rinse thoroughly hang and allow to dry 4 Reconnect the air tubing to the...

Страница 26: ...sterilization If you do not have Internet access please contact your ResMed representative Air tubing Replace the air tubing Alternatively consult the air tubing instructions for cleaning and disinfec...

Страница 27: ...e air tubing has been properly attached to the device and patient interface and humidifier if used Notes The alarm log and alarm settings are maintained when the device is powered down and in the even...

Страница 28: ...red stop treatment 4 Re start treatment LCD High Temperature 10 11 12 23 The temperature inside the device is too high Treatment may lead to stop Ensure the ambient temperature is within the specified...

Страница 29: ...exceeded the pre set alarm level Breathe normally to disable the alarm If the problem persists contact your clinician LCD Internal Battery Low The internal battery capacity is below 30 Connect the dev...

Страница 30: ...f you want to run from mains power Note The alarm will be cleared automatically after one minute Problem possible cause Solution No display Power failure The device stops delivering air pressure Remov...

Страница 31: ...s not deep enough to trigger SmartStart Take a deep breath in and out through the mask There is excessive leak Adjust position of mask and headgear Plug s missing from port s on mask Replace plug s Ai...

Страница 32: ...to the mains The humidifier does not work Return the device and the humidifier for servicing The water chamber is empty Fill the water chamber of the humidifier USB stick is not readable or writeable...

Страница 33: ...tion cap Therapy pressure tolerance IPAP 0 5 cm H2O 10 of set pressure end of inspiration EPAP PEEP 0 5 cm H2O 4 of set pressure CPAP 0 5 cm H2O 10 of set pressure Test condition T mode IPAP 40 cm H2O...

Страница 34: ...ompatibility Product complies with all applicable electromagnetic compatibility requirements EMC according to IEC60601 1 2 for residential commercial and light industry environments For further detail...

Страница 35: ...declaration electromagnetic immunity The device is intended for use in the electromagnetic environment specified below The customer or the user of the device should assure that it is used in such an...

Страница 36: ...nsidered If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above the device should be observed to verify normal operation If abnorma...

Страница 37: ...eled bin symbol indicates that the product bearing this symbol may not be disposed of together with general household waste but instead requires separate disposal This requirement for separate disposa...

Страница 38: ...ackage If the packaging is damaged the respective product must not be used and should be disposed along with the packaging Before using the device and the accessories for the first time ensure that al...

Страница 39: ...peration before changing parts Notes A note advises to special product features The above are general warnings and cautions Further specific warnings cautions and notes appear next to the relevant ins...

Страница 40: ...ability or fitness for a particular purpose Some regions or states do not allow limitations on how long an implied warranty lasts so the above limitation may not apply to you ResMed shall not be respo...

Страница 41: ...Spectrum Center Blvd San Diego CA 92123 USA ResMed UK Ltd 96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK See Resmed com for other ResMed locations worldwide ResMed SlimLine SmartStart St...

Отзывы: