background image

12

Insufficient air delivered from S8 Lightweight II

The following message is displayed on the LCD after you try to update settings 

or copy data to the Data Card: 

Card Error Remove Card

The following message is displayed on the LCD after you try to update settings 

or copy data to the Data Card: 

Settings Invalid Remove Card

The following message is displayed on the LCD after you try to update settings 

or copy data to the Data Card: 

Settings Error Remove Card

The following message is NOT displayed on the LCD after you try to update the 

settings using the Data Card: 

Settings Success Remove Card

Ramp Time is in use.

Wait for air pressure to build up or change ramp time.

Air filter is dirty.

Replace air filter. 

Air tubing is kinked or 
punctured.

Straighten or replace tubing.

Air tubing is not connected 
properly.

Check air tubing. 

Mask and headgear are not 
positioned correctly.

Adjust position of mask and headgear.

Cap(s) are missing from 
access port on mask.

Replace cap(s).

Pressure required for 
treatment may have changed.

See your clinician to adjust the pressure.

Data Card is not inserted 
correctly.

Check that you are inserting the Data Card into the Data 
Card slot with

 

the arrow facing up, while the S8 

Lightweight II is switched on and not delivering therapy 
(standby mode).

The Data Card is only partly 
inserted.

Ensure that the Data Card is inserted into the Data Card 
slot as far as it will go.

You may have removed the 
Data Card before settings 
were copied to the 
S8 Lightweight II.

Reinsert the Data Card and wait for the 

Settings

 

Success Remove Card

 message to appear on the 

LCD.

The card is not a ResScan Data 
Card.

Only use a ResScan Data Card.

The identification details on 
the Data Card do not match 
the identification details on 
your device.

Contact your clinician immediately.

There is a data error on the 
Data Card.

Contact your clinician immediately.

The settings were not 
updated.

Contact your clinician immediately.

Содержание Positive Airway Pressure Device S8

Страница 1: ...or contact details Protected by patents US 5199424 US 5522382 US 6705315 Other patents pending Protected by design registrations AU 302180 AU 302181 AU 302182 AU 302183 AU 302184 AU 302185 AU 302186 C...

Страница 2: ...weight II 3 Start Treatment 4 Stop Treatment 4 How to Use the Control Panel 4 How to Change the Settings on your S8 Lightweight II 6 Using the Data Card to Collect data 8 Cleaning and Maintenance 8 Da...

Страница 3: ...e airway pressure therapy may be contraindicated in some patients with the following pre existing conditions severe bullous lung disease pneumothorax pathologically low blood pressure dehydration cere...

Страница 4: ...page under Service Support WARNING Only ResMed mask systems are compatible for use with the S8 Lightweight II Please refer to Adverse Effects on page 1 Humidifiers A list of humidifiers compatible wi...

Страница 5: ...place the device where it can be bumped or where someone is likely to trip over the power cord If you put the device on the floor make sure the area is free from dust and clear of bedding clothes or...

Страница 6: ...s 5 Fit your mask as described in the mask user instructions WARNING A mask should not be used unless the S8 Lightweight II device is turned on and operating properly When ramping begins the LCD will...

Страница 7: ...htweight II is on Key Function Start Stop Starts or stops treatment Up Allows you to increase settings options and scroll through the S8 Lightweight II menu Down Allows you to decrease settings option...

Страница 8: ...time set by your clinician The standby screen displays the current ramp time in the right corner 1 Press or until the desired ramp time appears 2 Once a ramp time is selected begin therapy by pressin...

Страница 9: ...shes when a message is displayed Your clinician can set any of the following reminders on your LCD Message Description Action INSERT CARD May appear if your device is Data Card enabled Insert your Dat...

Страница 10: ...e card in its protective folder to your clinician using a postal envelope Using the Data Card to Update Settings If your clinician has provided a Data Card with new device settings 1 With the device i...

Страница 11: ...next use reconnect the air tubing to the air outlet and mask Monthly 1 Clean the exterior of the S8 Lightweight II with a damp cloth and mild liquid soap 2 Check the air filter for holes and blockage...

Страница 12: ...he S8 Lightweight II When you first use your S8 Lightweight II you may feel uncomfortable breathing against the airflow This is normal You will continue to breathe normally while you sleep The EPR set...

Страница 13: ...l adjustment is necessary but you will require an approved power cord for the region in which you are using the device Please consult the medical services department of your carrier if you intend to u...

Страница 14: ...ioned correctly Adjust position of mask and headgear Cap s are missing from access port on mask Replace cap s Pressure required for treatment may have changed See your clinician to adjust the pressure...

Страница 15: ...ION VALUES in accordance with ISO 4871 Sound Pressure Level 26 dBA with uncertainty of 2 dBA as measured according to ISO 17510 1 2002 28 dBA with uncertainty of 2 dBA as measured according to ISO 175...

Страница 16: ...is guide should not supersede instructions given by the prescribing physician A patient should not connect a device to the data communication port unless instructed to do so by their health care provi...

Страница 17: ...At a fixed rate of supplemental oxygen flow the inhaled oxygen concentration will vary depending on where the oxygen is introduced the pressure settings patient breathing pattern mask selection and l...

Страница 18: ...efective product must be made by the initial consumer at the point of purchase This warranty is in lieu of all other express or implied warranties including any implied warranty of merchantability or...

Страница 19: ...17 Limited Warranty...

Страница 20: ...18...

Отзывы: