How to Change the Settings on your S8 Lightweight II

You can change settings (such as Ramp Time) on the S8 Lightweight II by using 
the keypad and LCD. When the S8 Lightweight II is in standby mode you can 
display a series of screens (a menu) on the LCD screen in order to view and 
change the settings for a particular function. 

Use the menu screens to change the ramp time  and language. You can also view 
details about the usage and software version of your S8 Lightweight II. 

How to change the ramp time

During ramp time, the pressure increases from a low pressure to the prescribed 
treatment pressure. You can set the ramp time in 5-minute intervals, from 0 
minutes (OFF) up to a maximum ramp time set by your clinician. The standby 
screen displays the current ramp time in the right corner. 

1. Press 

 or 

 until the desired ramp time appears.

2. Once a ramp time is selected, begin therapy by pressing 

.

When ramping begins the LCD will display the word "RAMPING", with the 
current pressure indicated as dashes. Each dash is approximately equivalent 
to 2 cm H

2

O. For example, six dashes represents approximately 12 cm H

2

pressure. 

How to change the EPR level 

EPR can be used to help ease any discomfort you may experience when 
breathing out during treatment. If your clinician has allowed access you can 
change the EPR level. There are four settings: OFF; 1 (lowest EPR); 2; and 3 
(highest EPR).

S8 ESCAPE II

S8 CLASSIC

RAMP            10min
Menu

S8 CLASSIC

PATIENT MENU

USED HRS:       960

                            Exit

USAGE:      120/180

                           Exit

SW:   SX123456789

                           Exit

LANG:       ENGLISH

Change               Exit

EPR LEVEL:  1
Change              Exit

Only displayed if 
clinician enabled

Содержание Positive Airway Pressure Device S8

Страница 1: ...or contact details Protected by patents US 5199424 US 5522382 US 6705315 Other patents pending Protected by design registrations AU 302180 AU 302181 AU 302182 AU 302183 AU 302184 AU 302185 AU 302186 C...

Страница 2: ...weight II 3 Start Treatment 4 Stop Treatment 4 How to Use the Control Panel 4 How to Change the Settings on your S8 Lightweight II 6 Using the Data Card to Collect data 8 Cleaning and Maintenance 8 Da...

Страница 3: ...e airway pressure therapy may be contraindicated in some patients with the following pre existing conditions severe bullous lung disease pneumothorax pathologically low blood pressure dehydration cere...

Страница 4: ...page under Service Support WARNING Only ResMed mask systems are compatible for use with the S8 Lightweight II Please refer to Adverse Effects on page 1 Humidifiers A list of humidifiers compatible wi...

Страница 5: ...place the device where it can be bumped or where someone is likely to trip over the power cord If you put the device on the floor make sure the area is free from dust and clear of bedding clothes or...

Страница 6: ...s 5 Fit your mask as described in the mask user instructions WARNING A mask should not be used unless the S8 Lightweight II device is turned on and operating properly When ramping begins the LCD will...

Страница 7: ...htweight II is on Key Function Start Stop Starts or stops treatment Up Allows you to increase settings options and scroll through the S8 Lightweight II menu Down Allows you to decrease settings option...

Страница 8: ...time set by your clinician The standby screen displays the current ramp time in the right corner 1 Press or until the desired ramp time appears 2 Once a ramp time is selected begin therapy by pressin...

Страница 9: ...shes when a message is displayed Your clinician can set any of the following reminders on your LCD Message Description Action INSERT CARD May appear if your device is Data Card enabled Insert your Dat...

Страница 10: ...e card in its protective folder to your clinician using a postal envelope Using the Data Card to Update Settings If your clinician has provided a Data Card with new device settings 1 With the device i...

Страница 11: ...next use reconnect the air tubing to the air outlet and mask Monthly 1 Clean the exterior of the S8 Lightweight II with a damp cloth and mild liquid soap 2 Check the air filter for holes and blockage...

Страница 12: ...he S8 Lightweight II When you first use your S8 Lightweight II you may feel uncomfortable breathing against the airflow This is normal You will continue to breathe normally while you sleep The EPR set...

Страница 13: ...l adjustment is necessary but you will require an approved power cord for the region in which you are using the device Please consult the medical services department of your carrier if you intend to u...

Страница 14: ...ioned correctly Adjust position of mask and headgear Cap s are missing from access port on mask Replace cap s Pressure required for treatment may have changed See your clinician to adjust the pressure...

Страница 15: ...ION VALUES in accordance with ISO 4871 Sound Pressure Level 26 dBA with uncertainty of 2 dBA as measured according to ISO 17510 1 2002 28 dBA with uncertainty of 2 dBA as measured according to ISO 175...

Страница 16: ...is guide should not supersede instructions given by the prescribing physician A patient should not connect a device to the data communication port unless instructed to do so by their health care provi...

Страница 17: ...At a fixed rate of supplemental oxygen flow the inhaled oxygen concentration will vary depending on where the oxygen is introduced the pressure settings patient breathing pattern mask selection and l...

Страница 18: ...efective product must be made by the initial consumer at the point of purchase This warranty is in lieu of all other express or implied warranties including any implied warranty of merchantability or...

Страница 19: ...17 Limited Warranty...

Страница 20: ...18...

Отзывы: