KrosFlo® TFF System
–
Auxiliary Scale ACSS-20K
User Guide
22 repligen.com 420-14871-001 Rev. 00
Table 10. Working mode settings
Accessibility
Save mode
Time interval
Lo threshold
Weighing
3.1.1.Acc
3.1.2.Snn
3.1.3.Int
3.1.4.Lo
Parts Counting
3.2.1.Acc
3.2.3.Snn
3.2.4.Int
3.2.5.Lo
+/- control
3.3.1.Acc
3.3.2.Snn
3.3.3.Int
3.3.4.Lo
Percent Weighing %
3.4.1.Acc
3.4.3.Snn
3.4.4.Int
3.4.5.Lo
Peak Hold
3.5.1.Acc
-
-
3.5.2.Lo
Totalizing
3.6.1.Acc
3.6.2.Snn
3.6.3.Int
3.6.4.Lo
Animal Weighing
3.7.1.Acc
-
-
3.7.3.Lo
11.2.1
Working mode accessibility
To enable/disable given working mode, press
key.
1.
Enter
<P3.Func>
menu and select a given working mode.
2.
Go to
<Acc>
function.
3.
Press
key, parameter values are displayed successively one by one, where:
YES
–
working mode enabled,
no
–
working mode disabled.
4.
Press
key to confirm, next return to home screen.
11.2.2
Save mode
Parameter allowing to set mode of sending data from the weighing device to a peripheral device.
1.
Enter
<P3.Func>
menu and select a given working mode.
2.
Go to
<Snn>
function.
3.
Press
key, parameter values are displayed successively one by one, where:
Table 11. Save mode options
Option
Description
StAb
Manual printout of stable weighing result. Upon pressing
key at the moment
when the result is unstable (no
pictogram displayed), the program first waits
for the stability condition to be met, only then the printout is carried out
nStAb
Manual printout of each weighing result. In case of unstable indication,
<?> sign is displayed in front of the 'mass frame'
rEPL
Automatic printout of the first stable weighing result above <Lo> threshold (to set
<Lo> threshold go to <Lo> parameter)
rEPLi
Automatic printout with time interval set in [min] (to set the interval go to
<Int> parameter)
4.
Press
key to confirm, next return to home screen.
11.2.3
Automatic printout interval
Parameter enabling to set frequency of an automatic printout. Printout interval is set in minutes
with 1 [min] accuracy within 1 [min]
–
1440 [min] range.