KrosFlo® TFF System
–
Auxiliary Scale ACSS-20K
User Guide
16 repligen.com 420-14871-001 Rev. 00
2.
Select respective entry and press
key, tare value edit box is displayed.
3.
Enter tare value, to do this press
to select digit to be edited and
to set digit value
(0 - 9).
4.
Press
key to confirm,
<tArE 0>
is displayed.
5.
Press
to return to home screen.
5.6.2
Selecting tare value from the scale memory
1.
Enter
<P2.rEAd / 2.7.tArn>
submenu, name of tare no. 1 from tare database is displayed
(
<tArE 0>
), to select a different record press
key.
2.
To select a tare value press
key.
3.
The set tare value is displayed on the screen with a minus sign,
Net
symbol is shown in the
upper-left corner of the screen.
Note:
the tare value acquired from the weighing device memory is not remembered upon the
weighing device restart.
5.7
Last digit
Function designed to disable display of the last weighing indication digit, this results with less
accurate measurement.
1.
Enter
<P2.rEAd / 2.8.LdiG>
submenu.
2.
Press
key, filter values are displayed successively one by one, where:
ALAS
–
Last digit
always on,
nEur
–
Last digit always off, and
uuSt
–
Last digit on only when the weighing
indications is stable.
3.
Set respective value and press
key to confirm, next go to the home screen.
6.
Communication
Communication between the scale and peripheral devices is established via the following ports:
RS232 (1), USB type A, USB type B. To set the ports go to
<P4.Conn>
submenu.
6.1
RS232 (1) port settings
1.
Enter
<P4.Conn / 4.1.rS1>
submenu.
2.
Submenu
<4.1.1.bAd>
sets the baud rate. Press
key, and rates are displayed
successively one by one:
2400, 4800, 9600, 19200, 38400, 57600, 115200 bit/s.
3.
Submenu
<4.1.1.bAd>
sets the baud rate. Press
key, and rates are displayed
successively one by one:
2400, 4800, 9600, 19200, 38400, 57600, 115200 bit/s.
4.
Set respective value and press
key to confirm, next go to the home screen.
6.2
USB (A) port
USB port of type A is intended for:
•
Operator’s parameters
export/import