Renishaw VIONiC RTLC20-S Скачать руководство пользователя страница 10

VIONiC RTLC20-S installation guide

8

Readhead mounting and alignment

Mounting brackets

The bracket must have a flat mounting surface and should provide adjustment to enable conformance to 
the installation tolerances, allow adjustment to the rideheight of the readhead, and be sufficiently stiff to 
prevent deflection or vibration of the readhead during operation.

Readhead set-up

Ensure that the scale, readhead optical window and mounting face are clean and free from obstructions.

NOTE:

 When cleaning the readhead and scale apply cleaning fluid sparingly; do not soak. 

To set nominal rideheight, place the green spacer with the aperture under the optical centre of the 

readhead to allow normal LED function during set‑up procedure. Adjust the readhead to achieve a flashing 
Green LED along the full axis of travel. The faster the flash rate, the closer it is to optimum set‑up. 

The 

optional Advanced Diagnostic Tool ADTi-100 (A-6195-0100) and ADT View software can be used to 
optimise signal strength in challenging installations. See 

www.renishaw.com/adt

 for more information.

NOTE:

 When re-installing the readhead, factory defaults should be restored 

(see page 9)

.

Roll  

0° ±0.5°

Yaw 

0° ±0.4°

Pitch 

0° ±1°

Rideheight 

2.1 ±0.15 mm

Green spacer

Readhead set-up LED status

Green 

flashing

Orange 

flashing

Red 

flashing

Readhead LED diagnostics

Mode

LED

Status

Installation mode

Green flashing

Good set‑up, maximise flash rate for optimum set‑up

Orange flashing

Poor set‑up, adjust readhead to obtain Green flashing LED

Red flashing

Poor set‑up, adjust readhead to obtain Green flashing LED

Calibration mode

Blue single flashing

Calibrating incremental signals

Blue double flashing

Calibrating reference mark

Normal operation

Blue

AGC on, optimum set-up

Green

AGC off, optimum set-up 

Red

Poor set‑up; signal may be too low for reliable operation

Blank flash

Reference mark detected (visible indication at speed 

< 100 mm/s only)

Alarm

4 red flashes

Low signal or over signal; system in error

Содержание VIONiC RTLC20-S

Страница 1: ...RSLM high accuracy linear encoder Installation guide M 6195 9108 03 B VIONiC RTLC20 S linear encoder system ...

Страница 2: ...le application 5 End covers 5 Reference mark selector and limit magnet installation 6 Datum clamp 6 VIONiC encoder system quick start guide 7 Readhead mounting and alignment 8 System calibration 9 Output signals 10 Speed 10 Electrical connections 11 Output specifications 11 General specifications 12 Scale technical specifications 12 Reference mark 12 Limit switches 12 ...

Страница 3: ...series encoder system data sheet Renishaw part no L 9517 9678 Advanced Diagnostic Tool ADTi 100 data sheet Renishaw part no L 9517 9699 Advanced Diagnostic Tool ADTi 100 and ADT View software quick start guide Renishaw part no M 6195 9321 and the Advanced Diagnostic Tool ADTi 100 and ADT View software user guide Renishaw part no M 6195 9413 These can be downloaded from our website www renishaw com...

Страница 4: ... self adhesive tape is on the outside of bend 95 relative humidity non condensing to IEC 60068 2 78 Humidity Storage 70 C 0 C Operating Storage and handling N heptane CH3 CH2 5 CH3 Propan 2 ol CH3 CHOHCH3 Scale and readhead Readhead only Acetone CH3 COCH3 Methylated Spirits Chlorinated Solvents ...

Страница 5: ...on P and Q limit switch sensor position Selected IN TRAC reference mark 0 08 Roll tol 0 5 13 5 4 25 4 15 10 8 75 Offset 2 75 0 5 0 25 29 Detail A Scale reading surface 5 Mounting faces Scale thickness 0 4 including adhesive P limit magnet Reference mark selector magnet Q limit magnet Optical centreline incremental and reference mark Yaw tol 0 4 Rideheight 2 1 0 15 Pitch tol 1 2 mounting holes M2 5...

Страница 6: ...035 5 8 START See page 5 FINISH See page 5 A C End cover alignment indicator 0 5 F 0 2 100 0 05 F F axis of motion F axis of motion Nominal Q limit trigger point Ra 3 2 Optional bolted reference mark selector or limit magnets 22 18 10 3 7 Ø2 2 1 85 18 5 1 0 8 3 5 Detail A 12 1 5 4 5 3 7 Datum clamp A 9585 0028 RTLC20 S scale Detail C Detail B 15 1 IN TRAC reference mark Bolted magnet type Part num...

Страница 7: ... RTLC20 S scale installation drawing on page 4 3 Thoroughly clean and degrease the substrate using recommended solvents see Storage and handling on page 2 Allow the substrate to dry before applying scale 4 Mount the scale applicator to the readhead mounting bracket using M2 5 screws Place the shim supplied with the readhead between the applicator and substrate to set the nominal height NOTE The sc...

Страница 8: ...ed by magnetic materials in close proximity In such cases they should be held in place using an additional fillet of epoxy glue or similar along the outer edge of the magnet assembly Optional bolted reference and limit magnets are available X The reference mark selector and limit actuator locations are correct for the readhead orientation shown X External magnetic fields greater than 6mT in the vi...

Страница 9: ...ate the calibration routine The LED will single flash Blue Move the readhead along the scale at slow speed 100 mm s without passing a reference mark until the LED starts double flashing Blue The system is now calibrated and ready for use Calibration values Automatic Gain Control AGC and Automatic Offset Control AOC status are stored in readhead non volatile memory at power down NOTE If calibration...

Страница 10: ...optional Advanced Diagnostic Tool ADTi 100 A 6195 0100 and ADT View software can be used to optimise signal strength in challenging installations See www renishaw com adt for more information NOTE When re installing the readhead factory defaults should be restored see page 9 Roll 0 0 5 Yaw 0 0 4 Pitch 0 1 Rideheight 2 1 0 15 mm Green spacer Readhead set up LED status Green flashing Orange flashing...

Страница 11: ...ased X The system automatically exits the calibration routine and is ready for operation X AGC is automatically switched on once calibration is complete To switch off AGC refer to the Enabling disabling AGC section X If the LED continues double flashing after repeatedly passing the chosen reference mark it is not being detected Ensure that the correct readhead configuration is being used Readheads...

Страница 12: ...imum speed m s Minimum edge separation ns 5 µm D 1 µm X 0 5 µm Z 0 2 µm W 0 1 µm Y 50 nm H 40 nm M 25 nm P 20 nm I 10 nm O 5 nm Q 2 5 nm R 50 12 12 12 7 25 3 63 1 81 1 45 0 906 0 725 0 363 0 181 0 091 25 3 40 12 12 12 5 80 2 90 1 45 1 16 0 725 0 580 0 290 0 145 0 073 31 8 25 12 12 9 06 3 63 1 81 0 906 0 725 0 453 0 363 0 181 0 091 0 045 51 2 20 12 12 8 06 3 22 1 61 0 806 0 645 0 403 0 322 0 161 0 ...

Страница 13: ... isolator Remote CAL operation CAL 0 V Remote operation of the CAL AGC is possible via CAL signal Output specifications Digital output signals Form Square wave differential line driver to EIA RS422A except limits P and Q Incremental 2 channels A and B in quadrature 90 phase shifted Signal period P Resolution S A B Z Reference Synchronised pulse Z duration as resolution Bi directionally repeatable ...

Страница 14: ...C Installation temperature 15 C to 35 C Datum fixing Loctite 435 with A 9585 0028 clamp CAUTION Renishaw encoder systems have been designed to the relevant EMC standards but must be correctly integrated to achieve EMC compliance In particular attention to shielding arrangements is essential For lengths 2 m FASTRACK with RTLC20 is recommended Refer to the VIONiC RTLC20 FASTRACK linear encoder syste...

Страница 15: ...hts reserved Renishaw reserves the right to change specifications without notice RENISHAW and the probe symbol used in the RENISHAW logo are registered trade marks of Renishaw plc in the United Kingdom and other countries apply innovation and names and designations of other Renishaw products and technologies are trade marks of Renishaw plc or its subsidiaries Loctite is a registered trade mark of ...

Отзывы: