VIONiC RTLC20/

FASTRACK

 installation guide

10

System calibration

NOTE:

 The functions described below can also be carried out using the optional ADTi-100 and ADT View 

software. See 

www.renishaw.com/adt

 for more information.

Ensure signal strength has been optimised along the full axis of travel, the LED will be flashing Green. Cycle 
the power to the readhead or connect the ‘Remote CAL’ output pin to 0 V for < 3 seconds. The readhead will 
then single flash Blue to indicate it is in calibration mode as detailed in 

‘Readhead mounting and alignment’ on 

page 9

. The readhead will only enter calibration mode if the LED is flashing Green.

Step 1 – Incremental signal calibration

 

X

Move the readhead along the axis at slow speed (< 100 mm/s or less than the readhead maximum 

speed, whichever is slower) ensuring it does not pass a reference mark, until the LED starts 

double‑flashing indicating the incremental signals are now calibrated and the new settings are stored in 

the readhead memory. 

 

X

The system is now ready for reference mark phasing. For systems without a reference mark, cycle the 

power to the readhead or connect the ‘Remote CAL’ output pin to 0 V for < 3 seconds to exit 

 

calibration mode.

 

X

If the system does not automatically enter the reference mark phasing stage (LED continues single 

flashing) the calibration of the incremental signals has failed. After ensuring failure is not due to 
overspeed (> 100 mm/s or exceeding the readhead maximum speed) exit the calibration routine, restore 

factory defaults as detailed below, and check the readhead installation and system cleanliness before 
repeating the calibration routine.

Step 2 – Reference mark phasing

 

X

Move the readhead back and forth over the selected reference mark until the LED stops flashing and 

remains solid Blue (or Green if AGC is disabled). The reference mark is now phased. 

 

X

The system automatically exits the calibration routine and is ready for operation.

 

X

AGC is automatically switched on once calibration is complete. To switch off AGC refer to the 
‘Enabling/disabling AGC’ section.

 

X

If the LED continues double‑flashing after repeatedly passing the chosen reference mark it is not 

 

being detected.  

 

-

Ensure that the correct readhead configuration is being used. Readheads can either output all 
reference marks or only output a reference mark where a reference selector magnet is fitted 

depending on the options chosen when ordering.

 

-

Check reference mark selector magnet is fitted in the correct location relative to readhead 

orientation 

(see ‘RTLC20/FASTRACK carrier installation drawing’ on page 4)

.

Calibration routine manual exit

 

X

To exit the calibration routine at any stage cycle the power to the readhead or connect the ‘Remote CAL’ 
output pin to 0 V for < 3 seconds. The LED will then stop flashing.

Restoring factory defaults

When re-installing the system, or in the case of continued calibration failure, factory defaults should  
be restored. 

To restore factory defaults:

 

X

Switch system off.

 

X

Obscure the readhead optical window (using the spacer supplied with the readhead ensuring the cut-out  
is NOT under the optical window) or connect the ‘Remote CAL’ output pin to 0 V.

 

X

Power the readhead.

 

X

Remove the spacer or, if using, the connection from the ‘Remote CAL’ output pin to 0 V.

 

X

The LED will start continuously flashing indicating factory defaults have been restored and the readhead  
is in installation mode (flashing set‑up LED).

 

X

Repeat 

‘Readhead set-up’

 procedure, 

see page 9

.

Enabling/disabling AGC

The AGC is switched on once the system has been calibrated indicated by a Blue LED. AGC can be 

manually switched off by connecting the ‘Remote CAL’ output pin to 0 V for > 3 seconds < 10 seconds. The 

LED will then be solid Green. 

LED

Settings stored

Blue single flashing

None, restore factory defaults and recalibrate

Blue double flashing

Incremental only

Blue (auto-complete)

Incremental and reference mark

Содержание Vionic RTLC20/FASTRACK

Страница 1: ...VIONiC RTLC20 FASTRACK linear encoder system Installation guide M 6195 9225 03 A...

Страница 2: ...installation 5 Reference mark selector and limit magnet installation 7 VIONiC encoder system quick start guide 8 Readhead mounting and alignment 9 System calibration 10 Output signals 11 Speed 11 Ele...

Страница 3: ...5386081 US7550710 CN101300463 EP1946048 JP5017275 US7624513 CN101310165 EP1957943 US7839296 CN102057256 EP2294363 JP5475759 JP5755299 KR1550483 US8141265 CN1314511 EP1469969 JP5002559 US8987633 US8466...

Страница 4: ...sive tape is on the outside of the bend radius Storage and handling 70 C 20 C 95 relative humidity non condensing to BS EN 60068 2 78 2013 Humidity Storage 70 C 0 C Operating Scale and readhead Readhe...

Страница 5: ...Static bend radius 35 23 11 5 6 min Set up LED A 0 6 29 Detail A 4 2 5 0 2 5 Reference mark selector magnet FASTRACK carrier thickness 0 4 including adhesive Forward direction of readhead relative to...

Страница 6: ...A 9653 0143 Dimensions as P limit Ra 3 2 Q limit magnet A 9653 0139 Dimensions as P limit 1 5 Optional end cover Pair A 9589 0058 18 5 B A Do not cut FASTRACK carrier over shaded areas 18 0 1 15 1 P...

Страница 7: ...RACK carrier or scale Ensure the block is in the correct orientation as shown Whilst holding the block in place in a smooth motion pull down the lever to cut through the FASTRACK carrier or scale Befo...

Страница 8: ...um The datum clamp fixes the RTLC20 scale rigidly to the substrate at the location chosen CAUTION The metrology of the system may be compromised if the datum clamp is not used The datum clamp does not...

Страница 9: ...elow and aligned with the outer edge of the FASTRACK carrier Limit magnets can be positioned at any user defined location along the FASTRACK carrier but the reference mark selector magnet should be po...

Страница 10: ...o initiate the calibration routine The LED will single flash Blue Move the readhead along the scale at slow speed 100 mm s without passing a reference mark until the LED starts double flashing Blue Th...

Страница 11: ...head mounting and alignment Mounting brackets The bracket must have a flat mounting surface and should provide adjustment to enable conformance to the installation tolerances allow adjustment to the r...

Страница 12: ...d X The system automatically exits the calibration routine and is ready for operation X AGC is automatically switched on once calibration is complete To switch off AGC refer to the Enabling disabling...

Страница 13: ...z Maximum speed m s Minimum edge separation ns 5 m D 1 m X 0 5 m Z 0 2 m W 0 1 m Y 50 nm H 40 nm M 25 nm P 20 nm I 10 nm O 5 nm Q 2 5 nm R 50 12 12 12 7 25 3 63 1 81 1 45 0 906 0 725 0 363 0 181 0 091...

Страница 14: ...opto isolator Remote CAL operation CAL 0 V Remote operation of the CAL AGC is possible via CAL signal Output specifications Digital output signals Form Square wave differential line driver to EIA RS42...

Страница 15: ...al specifications FASTRACK technical specifications Form H W 0 2 mm 8 mm Pitch 20 m Accuracy at 20 C includes slope and linearity 5 m m calibration traceable to International Standards Linearity 2 5 m...

Страница 16: ...enishaw reserves the right to change specifications without notice RENISHAW and the probe symbol used in the RENISHAW logo are registered trade marks of Renishaw plc in the United Kingdom and other co...

Отзывы: