Readhead mounting and alignment 

Mounting brackets

The bracket must have a fl at mounting surface, ensure conformance to the installation tolerances, 
allow adjustment to the rideheight of the readhead, and be suffi ciently stiff to prevent defl ection or 
vibration of the readhead during operation. 

Readhead set-up

Ensure that the scale surface, readhead optical 
window and mounting face are clean and free 
from obstructions. To set nominal rideheight, 
position readhead spacer with the ‘ L’ shaped 
aperture under the optical centre to allow 
normal LED function during set-up procedure. 
Adjust the readhead to maximise the signal 
strength. On RGH20 direct output you 
should achieve a Green LED on the 
readhead over the full rotation. 
For RGH20F/REF systems the SETUP 
LED on the interface should be at least 
Green, ideally Blue or Purple over the 
full rotation. 

NOTE: 

When using REF system ensure 

that Automatic Gain Control 
(AGC) is switched off. 
(CAL/AGC LED on the interface is off).

NOTE:

 Ensure readhead 

fi xing screws are tightened 
to 0.5 Nm  to 0.7 Nm.

Limit switch

A limit switch signal is output when the readhead sensor passes over the magnetic actuator. 

Reference mark set-up

To ensure unidirectional repeatability, the reference mark requires phasing with the scale in the 
direction of normal datuming operation. A reference pulse is output in both directions, but repeatability 
is guaranteed only in the phased direction. Ensure readhead is set-up correctly with a Green LED 
indication over the full rotation and that the reference mark actuator is fitted correctly.

NOTE: 

It is recommended that a datum procedure is performed as part of any power-up sequence to 

ensure the correct datum position is recorded.

Phasing procedure

The readhead must be moved over the reference mark in the direction to be used for the datuming 
operation. The reference mark is phased correctly when the set-up LED on the readhead or interface 
fl ashes red for 0.25 seconds. If it fl ashes Orange or goes Blank, the reference mark adjuster screw 
should be turned anti-clockwise by 

1

 

8

turn and the procedure repeated until a red fl ash is obtained.

Orange Blank

Red

RGH20 

LED colour, except at 
reference mark position.

RGH20 

LED colour, except at 
reference mark position.

Green Orange   Red

   Yaw
0°±0.5°

 Roll
0°±1°

  Pitch
0°±1°

0.8 ±0.08 mm

Rideheight

RGH20F/REF 

SETUP LED colour, except at reference mark position.

Red   Orange  Blank

NOTE: 

Reference mark fl ashes only 

occur up to 100 mm/s traverse speed.

RGH20F/REF 

SETUP LED colour, except at 
reference mark position.

Green Orange   Red
Blue
Purple

NOTE:

 The RGH20F readhead does not have an LED.

NOTE:

 The RGH20F readhead does not have an LED.

Содержание RGH20

Страница 1: ...RGH20 RESR angle encoder system Installation guide M 9561 0005 06 A...

Страница 2: ...e EC Declaration of Conformity is available on request FCC compliance This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not caus...

Страница 3: ...Storage and handling N heptane CH3 CH2 5CH3 Propan 2 ol CH3CHOHCH3 Ring and readhead Acetone CH3COCH3 Methylated Spirits Chlorinated Solvents Readhead only Acetone CH3COCH3 Methylated Spirits Chlorina...

Страница 4: ...4 8 0 8 0 08 R20 dynamic bend radius R10 static bend radius 4 2 0 2 Reference mark sensor limit switch sensor position Ring and readhead optical centreline A 9559 0650 9 75 4 1 1 5 11 6 10 5 3 Limit...

Страница 5: ...40 75 30 100 30 100 20 103 20 103 00 104 40 104 20 114 70 114 50 150 40 150 20 200 40 200 20 206 50 206 10 208 80 208 40 229 40 229 00 254 80 254 40 300 40 300 20 350 40 350 20 412 70 412 30 417 40 4...

Страница 6: ...m pitch Line count DI mm 55 04 55 00 80 04 80 00 130 04 130 00 180 04 180 00 75 100 150 200 11 840 15 744 23 600 31 488 5 920 7 872 11 800 15 744 N holes equally spaced on PCD DH 3 5 through N holes e...

Страница 7: ...he radial displacement between the screw locations Adjust the screws so the radial displacement is the same at each screw location When adjusting identify the screw location with the lowest radial dis...

Страница 8: ...mm DS mm 30 028 30 015 37 028 37 015 55 039 55 020 80 045 80 023 80 045 80 023 80 045 80 023 95 045 95 023 130 052 130 027 52 57 75 100 103 104 115 150 DO mm DS mm 180 052 180 027 186 060 186 031 186...

Страница 9: ...ection used for the datuming operation If fitted ensure correct reference mark magnet is fitted and positioned Adjust the phasing of the reference mark to give a red flash on the readhead LED as the r...

Страница 10: ...rence mark requires phasing with the scale in the direction of normal datuming operation A reference pulse is output in both directions but repeatability is guaranteed only in the phased direction Ens...

Страница 11: ...diameter 4 2 0 2 mm Flex life 20 x 10 6 cycles at 20 mm bend radius The RGH20 readheads have been designed to the relevant EMC standards but must be correctly integrated to achieve EMC compliance In p...

Страница 12: ...eiver termination for 3 state output Analogue output type RGH20B Vq V0 V1 V2 Vq V0 V1 V2 120R Readhead A B Z E Q Cable Z0 120R Customer electronics 120R A B Z E Q Speed Digital readheads The tables be...

Страница 13: ...magnet NOTE RGH20 readheads are available with reference mark or limit switch output Select output option at order 3 state alarm For RGH20D X Z readheads only differentially transmitted signals forced...

Страница 14: ...repeat the calibration procedure Restoring factory defaults When re installing the system or in the case of calibration failure the factory defaults should be restored To restore factory defaults Swit...

Страница 15: ...9 g Interface 100 g Cable standard 34 g m UHV 23 g m Readhead cable Standard Double shielded outside diameter 4 2 0 2 mm Flex life 20 x10 6 cycles at 20 mm bend radius UHV Single tinned copper braided...

Страница 16: ...ne receiver circuitry Recommended signal termination Customer electronics 0 V 1k Interface A B Z Q E Cable Z0 120R 1k 120R 5 V A B Z Q E Analogue output REF0000 only Vq V0 V1 V2 Vq V0 V1 V2 120R Digit...

Страница 17: ...125 0 422 40 0 2 5 000 5 000 3 240 2 700 1 800 1 620 1 296 0 900 0 810 0 450 0 169 100 0 1 3 24 2 700 1 620 1 350 0 900 0 810 0 648 0 450 0 405 0 225 0 084 200 50 nm 1 620 1 350 0 810 0 675 0 450 0 4...

Страница 18: ...Vq Vq NOTE RGH20F readheads and REF interfaces are available with reference mark or limit switch output Select output option at order Voltage at VX X 1 100 0 0 Signal level Set up signal voltage propo...

Страница 19: ...ishaw com M 9561 0005 06 For worldwide contact details please visit our main website at www renishaw com contact RENISHAW and the probe emblem used in the RENISHAW logo are registered trademarks of Re...

Отзывы: