QUANTiC RTLC40-S installation guide

RTLC40-S scale specifications

Form

 

(H × W)

0.4 mm × 8 mm (including adhesive)

Pitch

40 μm

Accuracy 

(at 20 °C)

(includes slope and linearity)

RTLC40-S ±15 µm/m calibration traceable to International Standards

RTLC40H-S ±5 µm/m calibration traceable to International Standards

Linearity

RTLC40-S ±5 µm/m achievable with 2 point error correction

RTLC40H-S ±2.5 µm/m achievable with 2 point error correction

Supplied length

20 mm to 10 m

*

 (> 10 m available on request)

Material

Hardened and tempered martensitic stainless steel fitted with a self-adhesive backing tape

Mass

12.9 g/m

Coefficient of thermal expansion 

(at 20 °C)

10.1 ±0.2 µm/m/°C

Installation temperature

15 °C to 35 °C

Datum fixing

Loctite 435 with A-9585-0028 clamp

Reference mark

Type

Customer selected 

IN-TRAC

 reference mark, directly embedded into incremental track. 

Bi-directional position repeatability

Selection

Single reference mark selection by selector magnet (A-9653-0143) customer positioned.

L ≤ 100 mm 

Single reference mark at scale centre

L > 100 mm Reference marks at 50 mm spacing (first reference mark 50 mm from scale end)

Repeatability

Unit of resolution repeatability (bi-directional) across full system rated speed and 
temperature ranges

Limit switches

Type

Magnetic actuators; with dimple triggers Q limit, without dimple triggers P limit 

(‘RTLC40-S scale installation drawing’, page 4)

Trigger point

The limit output is nominally asserted when the readhead limit switch sensor passes the 

limit magnet leading edge, but can trigger up to 3 mm before that edge

Mounting

Customer placed at desired locations

Repeatability

<  0.1 mm

For lengths > 2 m 

FASTRACK 

 carrier with RTLC40 is recommended. Refer to 

QUANTiC 

 RTLC40 / FASTRACK 

 incremental linear encoder system

 installation guide 

(Renishaw part no. M-9417-9248).

Содержание QUANTiC RTLC40-S

Страница 1: ...Installation guide M 9417 9255 03 B QUANTiC RTLC40 S incremental linear encoder system...

Страница 2: ...8 Reference mark selector and limit magnet installation 9 QUANTiC quick start guide 10 Readhead mounting and alignment 11 System calibration 12 Restoring factory defaults 13 Switching Automatic Gain C...

Страница 3: ...ice may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation The user is cautioned that any changes or modif...

Страница 4: ...e self adhesive tape is on the outside of bend 95 relative humidity non condensing to IEC 60068 2 78 Humidity Storage 70 C 0 C Operating Storage and handling N heptane CH3 CH2 5CH3 Propan 2 ol CH3CHOH...

Страница 5: ...5 2 mounting holes M2 5 through counterbored 3 2 3 deep both sides NOTE The recommended thread engagement is 5 min 7 5 including counterbore and the recommended tightening torque is between 0 25 and...

Страница 6: ...028 15 1 10 15 1 5 Dimension from substrate surface Supplied with 2 M2 4 screws NOTES The reference mark selector and limit actuator locations are correct for the readhead orientation shown External m...

Страница 7: ...cator A 9589 0115 u 2 M2 5 screws Optional parts u End cover kit A 9585 0035 u Renishaw scale wipes A 9523 4040 u Lint free cloth u Loctite 435 dispensing tip P TL50 0209 u Guillotine A 9589 0071 or s...

Страница 8: ...h the guillotine as shown and place guillotine press block down onto the scale Ensure the block is in the correct orientation as shown Whilst holding the block in place in a smooth motion pull down th...

Страница 9: ...to be inserted through the applicator as shown below 6 Begin to remove the backing paper from the scale and insert scale into the applicator up to the start point as shown Ensure backing paper is rou...

Страница 10: ...hesive Loctite 435 in the cut out on the datum clamp ensuring none of the adhesive wicks onto the scale surface Dispensing tips are available 1 Remove the backing paper from the datum clamp Place the...

Страница 11: ...it trigger point The limit output is nominally asserted when the readhead limit switch sensor passes the limit magnet leading edge but can trigger up to 3 mm before that edge RTLC40 S scale installati...

Страница 12: ...at power down NOTE If calibration fails LED remains single flashing blue restore factory defaults by obscuring the readhead optical window on power up page 13 Repeat the installation and calibration...

Страница 13: ...sed to optimise signal strength in challenging installations See www renishaw com adt for more information NOTE When re installing the readhead factory defaults should be restored page 13 Roll 0 0 8 Y...

Страница 14: ...100 mm s or exceeding the readhead maximum speed exit the calibration routine restore factory defaults as detailed below and check the readhead installation and system cleanliness before repeating the...

Страница 15: ...t the Remote CAL output pin to 0 V u Power the readhead u Remove the spacer or if using the connection from the Remote CAL output pin to 0 V u The LED will start continuously flashing indicating facto...

Страница 16: ...em page 12 Can t initiate the calibration routine Signal size is 70 u Realign readhead to obtain a green flashing LED During calibration the LED on the readhead remains single flashing Blue even after...

Страница 17: ...been calibrated for referencing your system as only this reference mark will be guaranteed to remain phased u Clean the scale and readhead optical window and check for scratches then recalibrate the s...

Страница 18: ...s P Pink 11 8 A 14 Q Black 10 6 B 13 Alarm E Orange 3 13 F 3 Remote CAL CAL Clear 9 1 5 C 4 Shield Screen Case Case Case Case Ferrule Digital outputs Function Signal Colour 15 way D type L 15 way D ty...

Страница 19: ...24 24 18 13 7 25 3 626 1 813 25 1 40 24 24 24 14 50 5 80 2 900 1 450 31 6 25 24 24 18 13 9 06 3 63 1 813 0 906 51 0 20 24 24 16 11 8 06 3 22 1 611 0 806 57 5 12 24 24 10 36 5 18 2 07 1 036 0 518 90 0...

Страница 20: ...V R Vp Vq P Q Select R so that maximum current does not exceed 20 mA Alternatively use a suitable relay or opto isolator Remote CAL operation CAL 0 V IMPORTANT The shield should be connected to the ma...

Страница 21: ...tion of the signal period is available Contact your local Renishaw representative for more information Digital output signals Form Square wave differential line driver to EIA RS422A except limits P an...

Страница 22: ...to 2000 Hz 3 axes Mass Readhead 9 g Cable 26 g m EMC compliance IEC 61326 1 Readhead cable Single shielded outside diameter 4 25 0 25 mm Flex life 20 106 cycles at 30 mm bend radius UL recognised comp...

Страница 23: ...nce mark directly embedded into incremental track Bi directional position repeatability Selection Single reference mark selection by selector magnet A 9653 0143 customer positioned L 100 mm Single ref...

Страница 24: ...0 1453 524901 E uk renishaw com www renishaw com For worldwide contact details visit www renishaw com contact Part no M 9417 9255 03 B Issued 02 2021 M 9417 9255 03 Renishaw plc Registered in England...

Отзывы: