4.9
Trig
g
er Logic
Master reset function
RMP400 features a master reset function to assist users who have mistakenly changed the probe
settings into an unintended state.
The application of the master reset function will clear all current probe settings and return the probe to
default settings.
The default settings are as follows:
•
Radio switch-on
•
Radio switch-off
•
Auto-reset on, enhanced trigger filter on 8 ms
•
Hibernation mode on 30 s
•
Multiple probe mode off
The default settings may not be representative of the required probe settings. Further configuration of
RMP400 may subsequently be necessary to achieve the required probe settings.
To reset the probe
1. Insert the batteries or, if they have already been installed, remove them for 5 seconds and then
refit them.
2. Following the LED check, immediately deflect the stylus and hold it deflected until eight red
flashes have been observed (if the battery power is low, each red flash will be followed by a blue
flash).
3. Keep the stylus deflected until the “
Switch-on method
” setting is displayed, then release it.
4. Hold the stylus deflected for 20 seconds. The status LEDs will flash yellow eight times. A
confirmation for master reset is required; if no action is taken, the probe will time out.
5. To confirm that a master reset is required, release the stylus and then hold the stylus deflected
again until the eight yellow flash sequence has ended. This action clears all probe settings and
returns the probe to default settings. Following an LED check, the RMP400 returns to Trigger
Logic and displays “
Switch-on method
”.
6. Further configuration using Trigger Logic may be necessary to achieve the required probe
settings.
1.
1
> 5 s
2
3
LED check