Resetting the probe
The probe will automatically reset to the armed state when the PI 7-3 is powered up or the TP7M is fitted to the probe head. It may sometimes
be necessary to perform a manual reset, for example after fitting a new stylus. Press the RESET button on the front panel for 2 seconds to
reset the probe to the armed (seated) state.
Indicator
Colour
Function
POWER ON
Green
Mains power in
TYPE STD
Green
Kinematic probe selected
TYPE TP7
Green
TP7M probe selected
PROBE SEATED
Green
ON - Probe armed (seated)
OFF - Probe triggered or no probe connected
PROBE DAMPED
Yellow
PDAMP asserted
CAUTION:
Pressing the reset button inhibits probe triggers. Prior to pressing the button the CMM must be stationary, in manual
mode, with the stylus clear of the workpiece.
NOTE:
The action of unlocking and locking the PH10 PLUS motorised head will perform the same function as the RESET button.
TP7 user's guide
www.renishaw.com
Issued 10 2019
15