background image

 

18 

UM Neya EN rev.0 28.02.2022 

 

Message 

“Motor Imbalance” 

 

The centrifuge is equipped by an automatic system for imbalance detection. 

If it is used with a load not balanced correctly, the machine will automatically stop and signals a fault to the operator 
with intermittent beep and message "Motor Imbalance" on flashing red screen. If this occurs, you should wait until 
the centrifuge stops and open the lid. Opening the lid the error message will disappear. 

Once you open the lid to 

check the correct load balancing as described at § 8.

 

 

 

Message 

“ERROR check imbalance switch”

 

Note: 

The  system  for  the  imbalance  detection  is  magnetic,  so  it  can  sometimes  happen  that  it  remains 

"closed" and then the centrifuge gives the following error message "ERROR Check imbalance switch". 

To eliminate the error is sufficient to slightly push for a couple of seconds the motor shaft towards the rear part 
of the centrifuge by directly pressing 

the rotor by the hand. If it’s 

necessary, repeat the operation. 

When you switch on the centrifuge for the first time, it may give the above error message. The situation is normal 
and is due to the transport of the machine with tabs that hold the motor shaft. Follow the steps above to clear the 
error. 
 

 

Setting of programs (if present) 

The follow instructions are valid for centrifuges NEYA 10/10R and NEYA 16/16R. To set centrifugation parameters 
for NEYA 8 please refer to paragraph 8. With closed lid and centrifuge in standby mode (see Figure 4), press shortly 

SETUP/ESC 

 

key, display goes into SETUP menu and word “PROGRAM” starts to blink (se

e Figure 10). 

 

 

 

 

 

 

Figure 10 

Press  shortly  the  knob 

  to  confirm  the  flashing  selection,  display  passes  into  the  choose  of  program  to  be 

modified menu (see Figure 11). 
 
 
 
 
 
 
 
 
 

Figure 11 

 

 

 

 

 

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x x

x x x

x x x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

x x

x

x

x

x

x

x

x

x

x

x

x

x

x

x x x

x x x

x

x

x

x

x

x

x

x

x

x

x

x

x

x

Содержание NEYA 10

Страница 1: ...riginal instructions NEYA 8 NEYA 10 NEYA 16 NEYA 10R NEYA 16R Manufacturer REMI ELEKTROTECHNIK LTD Survey 65 1 Valiv Village Vasai East Palghar 401 208 India Editing by the Authorized Representative G...

Страница 2: ...e 9 5 Unpacking of the centrifuge 9 Content of package 9 First use 9 Getting started 9 Parts of the instrument 10 6 Technical specifications 11 7 Display and control panel 12 8 Operation 14 Switching...

Страница 3: ...tion if present 22 SPIN function if present 22 9 SYSTEM Menu 23 END cycle beep 23 Speed limit 23 Time Date adjustment 23 Temperature unit 23 Factory reset 23 10 Cleaning and maintenance 24 Greasing of...

Страница 4: ...ON for a hazardous medium risk situation which could lead to serious injury or death if not avoided ATTENTION for a hazardous situation with reduced risk which if not avoided can result in material da...

Страница 5: ...ent The centrifuge must only be used with the rotors and accessories supplied If you need to replace or purchase them contact your local distributor The load must always be balanced and symmetrically...

Страница 6: ...ces When using chemicals or solvents follow the manufacturer s safety data sheets In the following cases it is possible that the protection provided for the centrifuge is compromised The responsibilit...

Страница 7: ...he machine When opening and closing the lid do not grasp the part between the lid and the centrifuge do not insert your fingers into the centrifuge closure and do not grasp the lid locking mechanism t...

Страница 8: ...ids are spilled into the chamber or into the rotor housing the centrifuge must be thoroughly and entirely cleaned by specialized personnel In case of contact with infectious liquids and pathogenic ger...

Страница 9: ...EN rev 0 28 02 2022 3 CE marking data The equipment is made in accordance with Directive 2006 42 EC and the relevant and applicable Community Directives at the time of its placing on the market fac si...

Страница 10: ...packaging parts and conserve the total packaging Note In case of shipment of the centrifuge to technical assistance the user is required to pack it in its original box to send it for repair service In...

Страница 11: ...bove such as the commands and the cooling part may be present or absent and different according to the centrifuge model Observation window Display Commands panel Signalisation LED bar Lid of centrifug...

Страница 12: ...Power 220 10 V 50 Hz 450 W Table 2 Features NEYA 10 NEYA 10R Maximum capacity 4 x 175 ml swing out 6 x 100 ml fixed angle Maximum speed 4 500 rpm swing out 6 000 rpm fixed angle Setting RPM Yes Setti...

Страница 13: ...tion levels 0 9 0 min 9 max Deceleration levels 0 9 0 no brake 1 min 9 max Temperature range 10 40 C 14 114 F Precool function Yes Display temperature Yes C and F Spin function Yes Programs 10 program...

Страница 14: ...02 2022 Figure 4 Standby display Figure 5 Working cycle Figure 6 Protected program Figure 7 Alarm signaling NOTE some parameters of the above screens are different depending on the model centrifuge F...

Страница 15: ...OFF The ON OFF button permits to switch on and switch of the centrifuge 8 Operation Switching on the centrifuge Turn on the centrifuge by the button ON OFF on the left side Frontal view The display sw...

Страница 16: ...he same time pull the opening handle Warning As specified on safety sticker the warranty and safety rules are not guaranteed if this sticker is broken The above sticker is not removable and not modifi...

Страница 17: ...damage caused by the use during centrifugation of this type of materials is not covered by warranty Moreover also the centrifugation of explosive materials or having dangerous chemical reactions is st...

Страница 18: ...g it is absolutely necessary to distribute the loads symmetrically around the axis of the rotor If the loads were not well balanced and the centrifuge went into unbalance the centrifuge would stop imm...

Страница 19: ...refer to paragraph 8 With closed lid and centrifuge in standby mode see Figure 4 press shortly SETUP ESC key display goes into SETUP menu and word PROGRAM starts to blink see Figure 10 Figure 10 Press...

Страница 20: ...press more times the SETUP ESC button to return to standby display see Figure 4 Note if the last parameter down deceleration for the ventilated centrifuges temperature for refrigerated centrifuges is...

Страница 21: ...x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x...

Страница 22: ...rd PROGRAM starts to blink see Figure 17 Figure 17 Press shortly the knob to confirm the blinking selection the display passes in the menu of program number selection see Figure 18 Figure 18 Program n...

Страница 23: ...x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x x...

Страница 24: ...beep Speed limit Time Date adjustment Temperature unit Factory reset END cycle beep This function permits to enable or disable the audible signal of the end of centrifugation By the knob select this p...

Страница 25: ...l accessories with a neutral detergent suitable for use with the materials sprayed on a soft damp cloth If in doubt contact the detergent manufacturer If there are any sample leaks in the rotor or acc...

Страница 26: ...art up or operation The Imbalance rotor error appears on the display Place the centrifuge on a fixed and stable surface Load the rotor in a balanced and symmetrical way If the anomaly persists contact...

Страница 27: ...Use only original spare parts 13 Disposal of electronic equipment This equipment is subject to regulations for electronic devices Dispose of in accordance with local regulations The illegal disposal o...

Отзывы: