4.  PROGRAMMING THE CHAIR

  The items described in Section 4. are built-in features of the 

chair. Only those users who wish to utilize these features need 

to be concerned with this Section. The chair will operate as 

previously designed if these features are not used. The chair 

can be programmed in the same fashion from either switch 

located on the chair back.

4.1.  Safety/Shutdown Feature

  The chair has a built-in Safety/Shutdown feature. This feature 

causes the chair to shutdown or become inoperable. This 

feature prevents unauthorized use of the chair. (See Figure 

2).

4.1.1  To ENTER Safety/Shutdown Mode

4.1.1.1  Depress and HOLD either STOP switch.

4.1.1.2    With  the  STOP  switch  depressed,  depress  the   

LOWER BASE switch on the same side of the chair back.

4.1.1.3  A two second beep will be heard to indicate the chair 

will NOT move if any additional switches are depressed.

4.1.1.4  The LED’s on the Membrane Switches will blink every 

three seconds while in Safety/Shutdown Mode.

4.1.2  To EXIT Safety/Shutdown Mode, repeat the 

above sequence.

4.1.2.1  Depress and HOLD either STOP switch.

4.1.2.2  With the STOP switch depressed, depress the LOWER 

BASE switch on the same side of the chair back.

4.1.2.3  A two second beep will be heard to indicate the chair 

is functional, and movement of the chair WILL OCCUR when 

another switch is depressed.

4.1.2.4  The LED’s will come back on when the chair becomes 

operational. 

 4.2.  Individual Programming of the RAISE TOP,  

RECLINE TOP, RAISE BASE, and LOWER 

BASE switch located on the Chair Back or 

Footswitch. (See Figure 2 & 3).

4.2.1    These  individual  switches  may  be  programmed  to 

function in two ways. If the movement of the chair STOPS 

when the switch is released, this is called the MOMENTARY 

MODE of operation. If the movement of the chair continues 

after the switch is released, this is called the MAINTAINING 

MODE  of  operation.  The  factory  designed  setting  for  the 

RAISE TOP and RECLINE TOP switches located on the chair 

top is MAINTAINING MODE. The factory designed setting for 

the RAISE BASE and LOWER BASE switches located on the 

chair top is MOMENTARY MODE. If changed, the new settings 

will remain permanent as long as main power remains applied 

to the chair and will apply to the similar function on the other 

side.

 4.2.2 To change these settings the chair must be placed in 

Setup Mode. Once in Setup Mode, depressing one of these 

switches will change its present setting from MOMENTARY 

to  MAINTAINING  or  from  MAINTAINING  to  MOMENTARY. 

Depressing a switch repeatedly will cause the setting to toggle 

between the two modes of operation.

4.2.3  To ENTER Setup Mode and change current 

settings

4.2.3.1  Depress and HOLD either STOP switch.

4.2.3.2  With the STOP switch depressed, depress the   AUTO 

switch on the same side of the chair back.

4.2.3.3  A one second beep will be heard to indicate you 

have ENTERED Setup Mode. The LED’s will blink on and off 

while in Setup Mode. 

4.2.3.4    Depress  any  of  the  following  switches  (including 

footswitch), one at a time, to toggle its current setting - RAISE 

TOP, RECLINE TOP, RAISE BASE, OR LOWER BASE. When 

the switch is depressed, a SHORT beep (0.25 seconds) will 

indicate the switch is now in MOMENTARY MODE of operation. 

A  LONG beep (1.05 seconds) will indicate the switch is now 

in MAINTAINING MODE of operation.

4.2.3.5  When finished programming the chair, the chair MUST 

be taken out of Setup Mode.

4.2.4  To EXIT Setup Mode, repeat steps 1 & 2 

above.

4.2.4.1  Depress and HOLD either STOP switch.

4.2.4.2  With the STOP switch depressed, depress the  AUTO 

switch on the same side of the chair back.

4.2.4.3   A one second beep will be heard to indicate the chair 

has EXITED Setup Mode. The LED’s will come back on. 

IN-FX920

12

Содержание FX920

Страница 1: ...INSTALLATION AND OPERATING INSTRUCTIONS OWNERS MANUAL MODEL FX920 LOW VOLTAGE CONTROL CHAIR...

Страница 2: ...ucts Inc 3535 Kings Mills Road Mason Ohio 45040 2303 1 800 735 0357 www reliance medical com SERIAL NUMBER For your future reference mark the serial number in the space provided QUALITY MANAGEMENT SYS...

Страница 3: ...No 26 Headrest 11 4 PROGRAMMING THE CHAIR 12 4 1 Safety Shutdown Feature 12 To ENTER Safety Shutdown Mode 12 To EXIT Safety Shutdown Mode 12 4 2 Individual Programming of the RAISE TOP RECLINE TOP RAI...

Страница 4: ...e presence of important operating and maintenance servicing instructions in this Installation and Operating Instructions REMARQUE Amplifieuneproc dure pratiqueoucondition DANGER AVERTISSEMENT ou PR CA...

Страница 5: ...ORDINARY DEGREE The degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE EQUIPMENT not suitable for use in the presence of a FLA...

Страница 6: ...nterference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devi...

Страница 7: ...nstrument Stand or from wall receptacle NOTE If this chair is purchased in conjunction with a Reliance instrument stand the chair control cable supplied will provide proper operation Refer to the inst...

Страница 8: ...ou want the motion of the Chair to CONTINUE after either of these switches is released then see Section 4 for programming instructions 3 2 Chair Lift 3 2 1 Raising Lowering the Chair Lift with Membran...

Страница 9: ...chaise CONTINUE lorsque l un ou l autre de ces interrupteurs est rel ch Si vous d sirez que le mouvement de la chaise S ARR TE apr s que l un ou l autre de ces interrupteurs est rel ch consultez la s...

Страница 10: ...1 This feature is employed as a simple way for the operator to lower the chair lift and raise the chair back after completion of an examination or treatment The automatic return system is activated b...

Страница 11: ...e identical and completely independent of each other 3 7 3 To raise an arm push rearward on the middle joint of the arm support link while raising upward on the arm A stop limits upward motion of the...

Страница 12: ...If the movement of the chair continues after the switch is released this is called the MAINTAINING MODE of operation The factory designed setting for the RAISE TOP and RECLINE TOP switches located on...

Страница 13: ...MAINTAINING operation If a mistake is made depress the same switch again to return that switch to its previous mode of operation If at any time the user becomes lost or confused during the programmin...

Страница 14: ...n seconds 4 5 3 2 The RAISE BASE switches that are to be programmed for Home Mode must be in Momentary Mode 4 5 3 3 DEPRESS and HOLD either STOP switch located on the chair back 4 5 3 4 With the stop...

Страница 15: ...ove and push the flat spring towards the center of the power module until a click is heard Click may not be heard in all cases The drawer will come forward slightly 4 Once the flap has been pushed tow...

Страница 16: ...FOR VOLTAGE 2 REPLACE SOLENOID COIL 3 CONTACT FACTORY CHAIR BASE WILL NOT RAISE PUMP MOTOR DOES NOT RUN 1 NO POWER TO PUMP MOTOR ASS Y 2 UP LIMIT SWITCH STUCK OPEN 3 PUMP MOTOR BURNED OUT 4 OTHER 1 CH...

Страница 17: ...ing and notify personnel in area of danger Wear rubber gloves during all cleaning activity Use caution in cleaning around buttons stitching and wooden or other decorative trim since these solvents cou...

Страница 18: ...ormez le distributeur ou l usine du mod le et du num ro de s rie de l ensemble Ces num ros sont de la plaque qui est situ e l arri re pr s de la base de l unit centrale 2 Indiquez la couleur des pi ce...

Страница 19: ...ee Figure 7 8 1761073 Upholstered Seat Assembly 2 Base Assembly See Figure 8 9 1764273 Apron Board Assembly 3 26 Headrest Assembly See Figure10 10 1760973 Upholstered Apron Assembly 4 1760488 Footstep...

Страница 20: ...0 37 41 40 39 38 42 1 43 11 5 50 8 12 3 44 9 14 35 36 6 51 13 47 29 26 23 22 49 48 24 43 15 4 44 17 18 19 31 46 20 48 33 25 2 27 34 10 28 16 21 7 45 52 16 2 52 2 FIGURE 7 CHAIR TOP ASSEMBLY IN FX920 2...

Страница 21: ...mbly 13 1524599 Lock Handle L H 39 1546299 Ribbon Cable 14 1765199 Armrest Hinge Pin 40 1548098 Membrane Switch Assembly 15 1313180 Pin Footrest 41 0298199 BHCS 5 16 18 x 1 16 0396299 Spring Washer 42...

Страница 22: ...FIGURE 8 CHAIR BASE ASSEMBLY 21 10 11 19 20 7 5 9 18 22 38 36 35 24 26 25 29 27 15 39 8 12 6 30 17 43 38 32 13 23 39 30 17 31 1 40 39 2 3 4 16 17 14 37 34 28 IN FX920 22...

Страница 23: ...asher 33 1994699 Power Line Filter 6A 12 2721092 Hose Assembly 34 2720392 Grey Base Cable Assembly 13 1506392 Pressure Tube Assembly 35 2720292 Black Base Cable Assembly 14 2025199 P C Board Assembly...

Страница 24: ...1 2 10 9 3 4 6 5 7 8 11 41 38 26 42 40 43 20 30 31 32 33 37 34 27 28 29 35 36 12 13 15 16 19 18 14 17 25 24 23 22 21 18 39 FIGURE 9 LIFT ASSEMBLY IN FX920 24...

Страница 25: ...92 Cylinder Assembly High 33 1012499 Valve Housing 0526792 Cylinder Assembly Low 34 0363299 Relief Valve Plunger 13 Spacer Cylinder in Item 12 35 0784699 Relief Valve Spring 14 Bushing in Item 12 36 1...

Страница 26: ...1 3 5 6 9 9 2 11 8 7 13 12 14 15 16 17 4 10 18 FIGURE 10 26 HEADREST ASSEMBLY IN FX920 26...

Страница 27: ...1427099 Spring Plunger 13 1538899 Spring 14 1454999 L H C S 10 24 x 3 4 Socket 15 1434799 Pivot Cap Pad 16 0467195 Bushing 188 I D x 312 O D 17 1456199 Nut Acorn 1 4 20 Low Crown 18 1425699 Screw Shou...

Страница 28: ...IN FX920 28 FIGURE 10 WIRING DIAGRAM FX920...

Страница 29: ...IN FX920 29...

Страница 30: ...IN FX920 30...

Страница 31: ...ordance with the appropriate instructions manual for the products Warranty will be voided if installation is performed by a non authorized individual 3 The products have not sustained breakage or othe...

Страница 32: ...mer Service 1 800 735 0358 Technical Service 513 398 3937 Http www reliance medical com P N 27158 REV E Copyright 2008 Printed in U S A Revised 10 2016 8672 MADE IN AMERICA EC REP Authorized European...