Note: Anytime the operator’s button is depressed twice within 1.5 seconds, (which 

may be caused by operator’s inadvertent pressure on the button) the software will 

interpret this as a demand for calibration. Then a sustained display of [CAL] will be 

seen which initiates the calibration sequence. The calibration sequence must be 

successfully completed before further measurements can be accomplished. 

Once activated, after [====] is displayed and a “beep” tone is heard, touch the 

TONO-PEN VET unit to the cornea lightly and briefly, then withdraw. Repeat 

several times. The corneal surface needs only to be momentarily contacted; 

indentation is not required and may lead to inaccurate readings.

A chirp will sound and a digital IOP measurement will be displayed each time 

a valid reading is obtained. The single horizontal bar at the bottom of the 

LCD, indicating statistical reliability, will not be displayed with each single IOP 

measurement.

After four (4) valid readings are obtained, a final beep will sound and the 

averaged measurement will appear on the LCD along with the single bar 

denoting statistical reliability. 

 

Note: If a single row of dashes [- - - -] appears on the LCD after the final beep, an 

insufficient number of valid readings was collected. If this occurs, repeat the 
patient examination procedure, starting at step 1.

 

To take another measurement, reactivate the TONO-PEN VET unit by 

pressing and releasing the activation switch as described in step 5. 

 

Note: If not reactivated within 25 seconds, the TONO-PEN VET unit will 

automatically turn off and the LCD will clear, placing the device in sleep mode, to 

conserve battery life. If  this occurs, activate the TONO-PEN VET by pressing the 
Operator’s Button once.

 

Replace the Ocu-Film Tip Cover before using the TONO-PEN VET unit on 

another patient and before storage.

Note: Failure to replace the Ocu-Film Tip Cover between patients may result 

in unreliable readings of IOP by allowing debris to enter the transducer 

assembly. 

 

Note: A reused Ocu-Film Tip Cover may leak. This may cause damage to the strain 

gauge and void the Warranty. 

10.

11.

1.

13.

14.

15.

Taking a Measurement

Содержание TONO-PEN VET

Страница 1: ...User s Guide TONO PEN VET Tonometer Rx Only...

Страница 2: ...ontents Introduction 4 Warnings and Cautions 6 Setup 8 Calibration 11 Taking a Measurement 13 Interpreting the LCD 16 Maintenance Storage 17 Troubleshooting 19 Guidance and manufacturer s declaration...

Страница 3: ...nstrument the waveform produced by each touch to the anesthetized corneal surface is analyzed and stored for a statistical comparison process Each single valid IOP reading is digitally displayed on th...

Страница 4: ...Page Figure 1 TONO PEN VET Components in Storage Case TONO PEN VET Tonometer OCU FILM Tip Covers Stylus Blade Ocu Cel XL Batteries Instruction Manual Introduction...

Страница 5: ...O ENSURE THAT CORRECT OPERATION OF THE INSTRUMENT IS MAINTAINED AND TO GUARANTEE THE SAFETY AND RELIABILITY OF THE INSTRUMENT ANY REPAIR OR SERVICE MUST BE PERFORMED BY REICHERT INC OR AN AUTHORIZED S...

Страница 6: ...ANTY CAUTION Check to be sure that the batteries are installed correctly Incorrect installation could cause severe damage to the electronics and void the Warranty CAUTION USE OR STORAGE OF THE TONO PE...

Страница 7: ...onents are missing or damaged Install the batteries in the battery compartment see page 9 Perform the Calibration Procedure see page 11 Slide a new Ocu Film Tip Cover over the TONO PEN VET unit s stai...

Страница 8: ...tion slowly push the stylus forward and pry the cover up to remove Using the other end of the stylus insert the tip into the holes on the underside of the body mold Insert the tip perpendicular into t...

Страница 9: ...Page 10 Setup Figure 4 Replacing Batteries...

Страница 10: ...etween the first and second button press must be between 0 5 and 1 5 seconds The TONO PEN will beep and display CAL Wait approximately 15 seconds for the TONO PEN to beep The display will change from...

Страница 11: ...eplace the batteries and repeat the calibration check Note If CAL appears followed by a row of four dashes the unit did not go into calibration mode To correct this depress the button quicker two time...

Страница 12: ...e page 11 Note Calibration should only be performed on the TONO PEN once daily unless otherwise indicated by the instrument The TONO PEN does not require calibration after every patient examination CA...

Страница 13: ...f test of the LCD Liquid Crystal Display If any of the LCD segments are not displayed the TONO PEN requires service If a momentary display of CAL is seen followed immediately by a single row of dashes...

Страница 14: ...eadings are obtained a final beep will sound and the averaged measurement will appear on the LCD along with the single bar denoting statistical reliability Note If a single row of dashes appears on th...

Страница 15: ...ument is activated Two rows of dashes followed by a beep indicates that the instrument is ready to take measurements CAL indicates that the instrument is in the process of calibration check or the ins...

Страница 16: ...n the event of unanticipated readings or if a Good calibration check cannot be obtained To clean the probe post Remove the Ocu Film Tip Cover Spray the compressed gas onto the probe tip for approximat...

Страница 17: ...ent is used to protect the sensitive electronics against intrusion and should be removed only by the manufacturer Note Removal of the Attachment Screw will void the warranty Specifications Dimensions...

Страница 18: ...ion or watch Instructional DVD Old or improperly applied Ocu Film Tip Cover Replace Ocu Film Tip Cover Batteries are low Replace batteries Need to check calibration Perform calibration check procedure...

Страница 19: ...tion in a typical commercial or hospital environment Guidance and manufacturer s declaration electromagnetic emissions The TONO PEN VET is intended for use in the electromagnetic environment specified...

Страница 20: ...rding to frequency of transmitter meters 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23...

Страница 21: ...ed by an electromagnetic site surveys should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80...