Printer Interface

The printer output is provided for connecting an external printer through either of the two RS232 

serial ports. Both are DB9 position female connectors. In order to communicate properly with 

a printer, a Null Modem cable must be used and should be connected while the instrument is 

turned off. Proper Baud rate selection of the RS232 serial port is also required, and should be 

set at 9600. 

In order to set the output text string for the printer, computer interface is required. The options 

that exist for the output are as follows, and can be arranged according to individual require-

ments.

a = Temperature

b = Index

c = Index-TC

d = Solids

e = Solids-TC

f = Current Mode

g = Channel

h = Time

i = Date

j = ID

k = Method

Once communications with a computer is successful, type “sf” and press the enter key on the 

computer key board. This will display any current, text output string. To edit the string type sf, 

then a space and enter the desired letter codes. Example: 

sf cagh

. With the example entry of 

cagh

 the instrument will print Index-TC, Temperature, Channel and Time.

Computer Interface

The Mark III includes two serial communications ports (Fig 11). Both are 

DB9 position female connectors that are located on the left side of the 

instrument. Data may be downloaded to a computer, and custom channel 

files can be uploaded to the refractometer or deleted through these ports. 

Regardless of which of the serial ports is chosen on the Mark III, communi-

cations with a computer is possible.

Any Windows® based computer can be used successfully when interfacing 

with the Mark III. Communication through software such as HyperTerminal 

or Tera Term is required. HyperTerminal is usually found on all Windows® 

based systems as a standard software option. Recommended communi-

cation settings of the serial port from the computer are: Baud rate 9600, 

8 Data Bits, No Parity, 1 Stop Bit and Xon/Xoff Flow Control. Please note, 

that the Baud rate of the instrument must match that of the computer.

To connect the instrument to a computer use a compatible serial cable. 

Connection must be made while the instrument is turned off. To set up an 

output text string reference Printer Output.

Baud Rate Selection

Press READ to choose 

other port

Press SELECT to exit

REAR

        FRONT

 

     1200

     2400 

     2400 

     4800 

     4800

     

9600

 

  

   9600

    19200 

   19200

    38400 

   38400

    57600 

   57600

Fig 11

Содержание ABBE MARK III

Страница 1: ...ABBE MARK III REFRACTOMETER USER GUIDE...

Страница 2: ...instrument with the prism surfaces and instrument clean and dry to prevent premature failure of the prisms and seals due to any harmful effects from sample residue The instrument is programmed to rea...

Страница 3: ...Overview 6 Calibration 7 Set Point Calibration 7 Span Point Calibration 7 Editing Calibration Points 8 Operational Parameters Sample Temperature Effects on Readings 8 Automatic Temperature Correction...

Страница 4: ...III power input connector or the unit s power cord Damage to the Mark III instrument and or injury to the op erator may occur Warning The Mark III should be used in strict accordance with the instruc...

Страница 5: ...47 mm x 47 mm Illumination Six high intensity LED s Prism Assembly Stainless steel assembly with glass prisms Range Channel 1 1 3000 1 7000 nD 0 95 Brix per ICUMSA Readability 0 0001 nD 0 1 Brix Preci...

Страница 6: ...Focusing Eyepiece rotate to attain sharp focus of the X pattern and the shadowline Menu Key LCD Display Up Down Keys Select Key Read Key Illumination and Sample Prisms Shadowline Adjustment Knob rota...

Страница 7: ...p shadowline In these cases applying the sample with a rubber spatula to squeeze out trapped bubbles or lifting the upper prism sev eral times to remove the trapped air is required The following are t...

Страница 8: ...ch as contact lens material that can fit between the prisms when closed Solid samples must be shaped and polished correctly in order to be read The size of solid samples should be no larger than 35mm...

Страница 9: ...tion points These points can be chosen to either calibrate across the entire reading range of the instrument or to closely match a specific range where samples are typically read The instrument does n...

Страница 10: ...y lower ing the sample temperature will raise the refractive index This change in refractive index due to a change in temperature will cause an appar ent change in concentration unless the reading is...

Страница 11: ...me Computer Interface The Mark III includes two serial communications ports Fig 11 Both are DB9 position female connectors that are located on the left side of the instrument Data may be downloaded to...

Страница 12: ...products using custom channels These channels represent scales that have been developed to measure a specific product Optional channels can be selected only after they have been uploaded from a PC to...

Страница 13: ...ids Percent concentration Brix Solids TC Percent concentration temperature corrected Brix An alternate method of changing modes exists in the reading screen once a sample has been measured All that is...

Страница 14: ...s using the keys Time The time may be set in either AM PM or 24 00 00 format Use the or keys to change the displayed value To advance to the next time seqment use the key and the key to exit Date The...

Страница 15: ...ider flexibility Any tubing that is highly rigid may apply excessive force on the prism preventing it from closing completely This condition can adversely affect reading accuracy Note If a water bath...

Страница 16: ...ff with a suitable solvent to avoid possibly scratching the prism Use caution when removing samples or drying the prism Care and Cleaning Preparing the prism surfaces for samples is essential if consi...

Страница 17: ...8152 993530 Accessories Catalog No 13981000 Printer 110 VAC Catalog No 13982000 Printer 220 VAC Catalog No 13107040 901 Null Modem Cable Catalog No 13981000 103 Paper 5 Rolls Catalog No WFAG20047 FUS...

Страница 18: ...SA Toll Free 888 849 8955 Phone 716 686 4500 Fax 716 686 4555 email ai reichert com www reichert com Reichert Instruments GmbH Hubertusstrasse 2 D 82229 Seefeld Germany Tel 49 8152 993530 Fax 49 8152...

Отзывы: