ROUTINE REPAIRS
Safety measures
Safety measures during routine repairs should comply with the requirements described in
paragraph 2 .1 .1 of the present Operating Manual .
All repairs shall be carried out with PC
power supply cables disconnected from the mains.
General Instructions
The list of possible problems and troubleshooting is given in Table 2 .
Table 2
Device
Problem
Cause
Solution
USB-Device for
MOV
When connecting
PC USB-port after
initialization the power
supply indicator does
not come on
USB connector is not
connected
Check the power cord, unplug
it and plug it back
The controller or power cord
has broken down
Service center repairs
Spontaneous magnetic
tape (MT) stopping in the
Optic-Mechanical Block
(OMB) tape drive tract
Polluting the tract by
foreign objects
Clean the polluted tract as
described in paragraph 4 .2
Straighten or replace the
tape
Magnetic tape jamming
due to its ruptures
Low stitching quality —
image doubling
Magnetic tape sliding
due to the tape drive
mechanism pollution
Clean the tape drive
mechanism as described in
paragraph 3 .1
The device does not
respond to the executive
commands
Software error
Restart the software;
if necessary restart the
operating system
Magnetic
Copying
Accessories Set
(MCAS)
Rubber sleeve gasket
of the Combination
Magnetic Scanner with
rotary concentrator
is damaged
Natural wear-and-tear
Replace the rubber sleeve
gasket as described in
paragraph 4 .2
No precision when
viewing the stitched
image
Improper preparation of
the investigated surface;
or a thick lacquer-and-
paint coating
Improve preparation of
the investigated surface,
measure the thickness of
lacquer-and-paint coating
When examining the
stitching image, there
are image fragments
irrelevant to the
investigated object
Magnetic tape is not fully
demagnetized
Demagnetize the magnetic
tape and make a new copy
When examining
the stitched image,
there are longitudinal
tracks irrelevant to the
investigated object
Adhesion of magnetic
particles to the scanner
surface or scratches on
the MT surface
Clean the scanner as
described in paragraph 3 .2,
demagnetize or replace the
MT and make a new copy
Instructions on Repair
Cleaning the tape drive of the Optic-Mechanical Block tract
Cleaning the tape drive of the Optic-Mechanical Block tract is done when it is polluted with
foreign objects, impeding the magnetic tape travel . The same is applied to the
MT extraction
,
when it was jammed during the automatic injection or reverse motion . Such jamming is
possible when the user does not observe the recommended edge form of the magnetic tape
strip (see paragraph 2 .2 .2 .2) . The OMB tape drive clearing and the MT extraction shall be
performed in the following order:
{
turn the front side of the device, when in the working position, to the operator
(see Figure 1),
{
release the pinch roller fork (see Figure 3, pos . 5) . Insert a thin core into an aperture
(Figure 11) and slightly press – until the lever of the roller is released,
4
4.3
4.2