cleaning/disinfection and sterilization, and that the validated 

parameters are adhered to during every cycle.

4.2 Please also observe the applicable legal requirements 

in your country as well as the hygiene regulations of the 

hospital or clinic. This applies especially with regard to the 

additional requirements for the inactivation of prions.

5. POST-OPERATIVE TREATMENT

The post-operative treatment must be carried out 

immediately, no later than 30 minutes after the completion 

of the operation. Additional information, while necessary, 

is provided in the respective product-specific usage 

instructions.

Steps:

1.Rinse away any surface soiling on the product with 

distilled deionized water or with a cleaning agent.

2.Rinse through all lumina (E.g. irrigation and aspiration 

connection) at least 3 times in the normal direction of flow 

(no back rinsing) using a disposable syringe (min. volume 

50 ml) filled with distilled/deionized water applied to the 

back nozzle.

3.An aldehyde-free cleaning and disinfection solution that 

is compatible with the products may also be used as an 

alternative rinsing solution. In such cases, it is necessary to 

rinse thoroughly afterwards at least 3 times with distilled, or 

deionized water.

6. CLEANING/DISINFECTION:

6.1 MECHANICAL REPROCESSING

Disinfection must be performed no later than 2 hours after 

the cleaning phase.

A machine cleaning and disinfection method should always 

be used for cleaning / disinfection because of the increased 

effectiveness of this method.

6.2 MECHANICAL CLEANING AND DISINFECTION

6.2.1 Thermal disinfection should be used if this function is 

available on your disinfector. Use if possible a disinfecting 

cycle compliant with the standard EN ISO 15883.

6.2.2 Note that there is a risk of disinfectant residue on 

products when using chemical disinfectants.

6.2.3 Ensure the following criteria are met when selecting a 

disinfector system:

•Disinfector is proven effective through testing (E.g. FDA 

approved or CE marked / EN ISO 15883 compliant).

•Disinfector has suitable baskets to hold small fragile 

products and has rinsing connections for the attachment to 

product lumina.

•The cleaning program is suitable for products to be 

processed and the rinsing cycle is sufficient.

•Only low microbe count (<10 cfu/ml) distilled or deionized 

water is used for all rinsing steps. (E.g. Aqua purificata, as 

per the specifications of Pharm. Eur. or USP).

•Air used for drying is HEPA filtered.

•Disinfector is serviced and checked on a regular basis.

Содержание MaxCure5

Страница 1: ...Please read this manual before operating RF RCL M014 Edition V2 0 Modify 20210529...

Страница 2: ...nting 8 4 Operation 9 5 Cleaning disinfecting and sterilizing 10 6 Transportation storage and maintenance 11 7 Troubleshooting 12 8 Environmental protection 14 9 Statement 14 10 EMC Declaration of con...

Страница 3: ...onductors 6 Please recharge the battery in cool and ventilated room 7 It is forbidden to self taking apart the battery in order not to result in short circuit or leakage 8 Keep the Curing Light clean...

Страница 4: ...NING No modification of this equipment is allowed WARNING If this equipment is modified appropriate inspection and testing must be conducted to ensure continued safe use of equipment WARNING If the cu...

Страница 5: ...against electric shock CLASS II EQUIPMENT According to the degree of protection against electric shock Type B applied part Screw inside outside Used indoor only Recovery Keep dry Handle with care Atmo...

Страница 6: ...should be used by trained qualified dentist This Curing Light is used for the principle of ray radiation to solidify the light sensitive resin by shooting at it in a short time it is used to restore t...

Страница 7: ...cal shock Class II 2 Protection degree against electrical shock Type B 3 Protection against harmful ingress of water or particular matter ordinary equipment IPX0 can t be waterproof 4 operation mode s...

Страница 8: ...art optical fiber Light guide Light Intensity 1000mW cm2 1800mW cm2 1000mW cm2 2500mW cm2 Modes setting TURBO mode Display P1 Illumination 1600mW cm2 1800mW cm2 NORMAL mode Display P2 Illumination 100...

Страница 9: ...LED lamp battery and power adapter Table 2 Technical Parameters of built in models Parameters MaxCureG MaxCureH Power supply Input 24VAC 50Hz 60Hz Input 24VAC 50Hz 60Hz Light Intensity 1000mW cm2 2500...

Страница 10: ...W 5W Composed mainly main unit LED lamp light hood cable main unit optical fiber light hood LED lamp cable 3 Installation and demounting 3 1 Installation of the table top models a Take off the red cap...

Страница 11: ...r is cut off The two power wire should be a little longer than the nylon thread to keep the power wire safe b Take off the red cap from the optical fiber and insert the metal part into the front of th...

Страница 12: ...operate the curing light for 10 times under 20s working mode then it will come into over heat protection status And only after 2 minutes sleep it can restart working 200s continuously 5 Cleaning disin...

Страница 13: ...Avoid solarization and getting wet in rain or snow during transportation 6 2 Storage 6 2 1 Don t store the machine together with the articles that is combustible poisonous caustic or explosive 6 2 2...

Страница 14: ...er 2 The power is off 1 Check the connection of the LED and the power 2 Make sure the power is on Light intensity insufficient 1 Optical fiber isn t inserted well to the bottom 2 The optical fiber has...

Страница 15: ...of the optical fiber 1 Reinstall the optical fiber 2 Change a new optical fiber 3 Clear the resin The equipment is not charging when the adapter is connected 1 The adapter is not connected well 2 Faul...

Страница 16: ...h EN 60601 1 2 for EMC This does not guarantee in any way that this device will not be effected by electromagnetic interference Avoid using the device in high electromagnetic environment Guidance and...

Страница 17: ...such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD lEC 61000 4 2 8 kV contact 2 kV 4 kV 8 kV 15 kV air 8 kV conta...

Страница 18: ...power quality should be that of a typical commercial or hospital environment If the user of the models Curing Light require continued operation during power mains interruptions it is recommended that...

Страница 19: ...calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 3 5 V1 P1 2 d 1 2 P1 2 80 MHz to 800 MHz d 2 3 P1 2 800 MHz to 2 5 GHz where P is the max...

Страница 20: ...F transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the models Curing Light are used exceeds the applicable RF compliance level a...

Страница 21: ...smitter W Separation distance according to frequency of transmitter m 150kHz to 80MHz d 1 2 P1 2 80MHz to 800MHz d 1 2 P1 2 800MHz to 2 5GHz d 2 3 P1 2 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2...

Страница 22: ...ts of the assembly are cleaned first 2 5 If you find that the reprocessing instructions from the manufacturer seem to be inadequate please inform manufacturer about those inadequacies 2 6 We encourage...

Страница 23: ...o rinse thoroughly afterwards at least 3 times with distilled or deionized water 6 CLEANING DISINFECTION 6 1 MECHANICAL REPROCESSING Disinfection must be performed no later than 2 hours after the clea...

Страница 24: ...examine the instruments visually 4 Carefully place the products in the disinfection basket Fastening of the products is only permissible if they are freely moveable in the fixture The products are no...

Страница 25: ...of other sterilization procedures E g ethylene oxide formaldehyde and low temperature plasma sterilization In such cases please observe the respective valid standards EN IS0 14937 ANSI AAMl IS0 14937...

Страница 26: ...td No 8 3 Information Industrial Park High Tech Zone Qixing District Guilin Guangxi 541004 P R China Tel 86 773 7796686 Email refine refine med com Website http www refine med com MedNet EC REP GmbH B...

Страница 27: ...ible for any troubles caused by quality problems or products technique and structure 3 The followings are beyond our warranty 1 The damage caused by disobeying the operation instruction or lack of the...

Страница 28: ...omer Address Post Code Tel E mail Purchase Date Production Date Distributor Model Product No Guilin Refine Medical Instrument Co Ltd No 8 3 Information Industrial Park High Tech Zone Qixing District G...

Отзывы: