17
Guidance & Declaration - Electromagnetic immunity
The models Curing Light are intended for use in the electromagnetic environment specified below. The customer or the
user of the models Curing Light should assure that it is used in such an environment.
Immunity test IEC 60601 test
level
Compliance
level
Electromagnetic environment - guidance
Conducted RF
lEC 61000-4-6
Radiated RF
lEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment should be used
no closer to any part of the models Curing Light, including cables, than
the recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d=[3,5/V1]×P
1/2
d=1.2×P
1/2
80 MHz to 800 MHz
d=2.3×P
1/2
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter In
watts (W) according to the transmitter manufacturer and d Is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance level
in each frequency range.
b
Interference may occur In the vicinity of equipment marked with the
following symbol:
NOTE 1 At 80 MHz end 800 MHz. the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.