Direct and Group Mode
This meter offers two operating modes: Direct and Group Mode. In the
Menu, choose “Options”, and select “Working mode” to select to work in
either Direct Mode or to select a Group.
Direct Mode is intended for quick and occasional readings. In this mode,
individual readings are logged to the memory provisionally. When the
meter is turned off or switched to GROUP Mode, all readings will be
deleted, but the statistic values won’t be changed until you take new mea-
surements. The readings and statistical values can be shown on the LCD
screen. The statistical analysis program can evaluate 80 readings. When
the memory is full, new readings will replace old readings. In this mode, it
has individual calibration values and limit values.
In Group Mode, there are four memory groupings that can store a maxi-
mum of 80 single readings, and 5 statistic values. Calibration values and
limit values can individually be set and stored for every group. When the
memory is full measuring will still continue, but will not be stored and
statistic values won’t be changed. You can delete group data and statistic
values, and reset the calibrate values and limit values. When measuring in
GROUP Mode, the symbol “GROx” will appear on the LCD screen. Press
the Power Button while holding the probe in the air, the meter will switch
to DIRECT mode, and the last reading will be shown (if available).
Calibration Methods
There are four different methods available for calibration:
Basic calibration
: Recommended for measurements on even surfaces
and if the measuring object has the same material, size, and curvature as
the zero plate attached in the casing.
Zero-point calibration
: Recommended if measuring errors up to +-(3%
of reading plus constant error of probe) are permitted (Example: Fe 1um;
no-Fe 1.5um).
One-point calibration (calibrating using calibration foil):
Recom-
mended if expected readings are close to the calibration value and if the
permitted error of probe is max. +-(1%...3% of reading plus constant
probe error).
continued ...